Application of the AMPLATZER Vascular Plug in Endovascular Occlusion of Dialysis Accesses

Journal Article · · Cardiovascular and Interventional Radiology
; ; ;  [1];  [2]
  1. Baskent University Faculty of Medicine, Department of Radiology (Turkey)
  2. Baskent University Faculty of Medicine, Department of General Surgery (Turkey)
The purpose of this study was to share our initial experience with the AMPLATZER Vascular Plug (AVP) in occluding dialysis accesses. Between January 2007 and October 2008, five patients with autogenous and one patient with prosthetic accesses were referred for endovascular occlusion owing to central venous obstruction (n = 4) and dialysis-associated steal syndrome (n = 2) leading to disabling complications. We used a single AVP in two patients, double AVPs in two patients, and double AVPs and n-butyl 2-cyanoacrylate in one patient with an autogenous access. Two coils were deployed between two AVPs to occlude the prosthetic graft. Immediate success was achieved in all patients, with no complications. Mean follow-up time was 13 months (range, 1-21 months) and none of the patients had experienced symptom recurrence as of the time of writing. The AVP allows simple, precise, and reliable dialysis access occlusion without significant complications when surgical ligation is not preferred. It can be used with other embolizing agents to provide occlusion in the case of failure.
OSTI ID:
21429112
Journal Information:
Cardiovascular and Interventional Radiology, Journal Name: Cardiovascular and Interventional Radiology Journal Issue: 5 Vol. 32; ISSN 0174-1551; ISSN CAIRDG
Country of Publication:
United States
Language:
English

Similar Records

Early Experience with the Amplatzer Vascular Plug II for Occlusive Purposes in Arteriovenous Hemodialysis Access
Journal Article · Sun Feb 14 23:00:00 EST 2010 · Cardiovascular and Interventional Radiology · OSTI ID:21429052

Transhepatic Preoperative Portal Vein Embolization Using the Amplatzer Vascular Plug: Report of Four Cases
Journal Article · Wed Nov 14 23:00:00 EST 2007 · Cardiovascular and Interventional Radiology · OSTI ID:21090737

Aneurysm of an Anomalous Systemic Artery Supplying the Normal Basal Segments of the Left Lower Lobe: Endovascular Treatment with the Amplatzer Vascular Plug II and Coils