Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Impact of Stent Design on In-Stent Stenosis in a Rabbit Iliac Artery Model

Journal Article · · Cardiovascular and Interventional Radiology
;  [1];  [2]; ;  [1];  [3];  [4]; ; ;  [1]
  1. University Hospital Heidelberg, Department of Diagnostic and Interventional Radiology (Germany)
  2. University Hospital Heidelberg, Department of General, Visceral and Transplantation Surgery (Germany)
  3. University Hospital Cologne, Department of Diagnostic Radiology (Germany)
  4. University Hospital Heidelberg, Department of Pathology (Germany)

The purpose of this study was to evaluate the impact of stent design on in-stent stenosis in rabbit iliac arteries. Four different types of stent were implanted in rabbit iliac arteries, being different in stent design (crown or wave) and strut thickness (50 or 100 {mu}m). Ten stents of each type were implanted. Each animal received one crown and one wave stent with the same strut thickness. Follow-up was either 12 weeks (n = 10 rabbits) or 24 weeks (n = 10 rabbits). Primary study end points were angiographic and microscopic in-stent stenosis. Secondary study end points were vessel injury, vascular inflammation, and stent endothelialization. Average stent diameter, relative stent overdilation, average and minimal luminal diameter, and relative average and maximum luminal loss were not significantly different. However, a trend to higher relative stent overdilation was recognized in crown stents compared to wave stents. A trend toward higher average and minimal luminal diameter and lower relative average and maximum luminal loss was recognized in crown stents compared to wave stents with a strut thickness of 100 {mu}m. Neointimal height, relative luminal area stenosis, injury score, inflammation score, and endothelialization score were not significantly different. However, a trend toward higher neointimal height was recognized in crown stents compared to wave stents with a strut thickness of 50 {mu}m and a follow-up of 24 weeks. In conclusion, in this study, crown stents seem to trigger neointima. However, the optimized radial force might equalize the theoretically higher tendency for restenosis in crown stents. In this context, also more favorable positive remodeling in crown stents could be important.

OSTI ID:
21428984
Journal Information:
Cardiovascular and Interventional Radiology, Journal Name: Cardiovascular and Interventional Radiology Journal Issue: 3 Vol. 33; ISSN 0174-1551; ISSN CAIRDG
Country of Publication:
United States
Language:
English

Similar Records

Carotid Artery Stenting Using a Novel Self-Expanding Braided Nickel-Titanium Stent: Feasibility and Safety Porcine Trial
Journal Article · Tue Sep 15 00:00:00 EDT 2009 · Cardiovascular and Interventional Radiology · OSTI ID:21429114

Self-Expanding Nitinol Renal Artery Stents: Comparison of Safety and Efficacy of Bare Versus Polyzene-F Nanocoated Stents in a Porcine Model
Journal Article · Mon Feb 14 23:00:00 EST 2011 · Cardiovascular and Interventional Radiology · OSTI ID:21608808

Evaluation of Silicone as an Endovascular Stent Membrane: In Vivo Canine Studies
Journal Article · Sat Sep 15 00:00:00 EDT 2001 · Cardiovascular and Interventional Radiology · OSTI ID:21083490