First Magnetic Resonance Imaging-Guided Aortic Stenting and Cava Filter Placement Using a Polyetheretherketone-Based Magnetic Resonance Imaging-Compatible Guidewire in Swine: Proof of Concept
Journal Article
·
· Cardiovascular and Interventional Radiology
- University Hospital Basel, Institute of Radiology (Switzerland)
- Kantonsspital Bruderholz, Institute of Radiology (Switzerland)
- Biotronik, Vascular Intervention (Switzerland)
- University of Duisburg-Essen, Department of Diagnostic and Interventional Radiology and Neuroradiology (Germany)
- University of Duisburg-Essen, Institute of Pathophysiology (Germany)
- University of Duisburg-Essen, Department of General, Visceral, and Transplantation Surgery (Germany)
- MagnaMedics GmbH (Germany)
The purpose of this study was to demonstrate feasibility of percutaneous transluminal aortic stenting and cava filter placement under magnetic resonance imaging (MRI) guidance exclusively using a polyetheretherketone (PEEK)-based MRI-compatible guidewire. Percutaneous transluminal aortic stenting and cava filter placement were performed in 3 domestic swine. Procedures were performed under MRI-guidance in an open-bore 1.5-T scanner. The applied 0.035-inch guidewire has a PEEK core reinforced by fibres, floppy tip, hydrophilic coating, and paramagnetic markings for passive visualization. Through an 11F sheath, the guidewire was advanced into the abdominal (swine 1) or thoracic aorta (swine 2), and the stents were deployed. The guidewire was advanced into the inferior vena cava (swine 3), and the cava filter was deployed. Postmortem autopsy was performed. Procedural success, guidewire visibility, pushability, and stent support were qualitatively assessed by consensus. Procedure times were documented. Guidewire guidance into the abdominal and thoracic aortas and the inferior vena cava was successful. Stent deployments were successful in the abdominal (swine 1) and thoracic (swine 2) segments of the descending aorta. Cava filter positioning and deployment was successful. Autopsy documented good stent and filter positioning. Guidewire visibility through applied markers was rated acceptable for aortic stenting and good for venous filter placement. Steerability, pushability, and device support were good. The PEEK-based guidewire allows either percutaneous MRI-guided aortic stenting in the thoracic and abdominal segments of the descending aorta and filter placement in the inferior vena cava with acceptable to good device visibility and offers good steerability, pushability, and device support.
- OSTI ID:
- 21428548
- Journal Information:
- Cardiovascular and Interventional Radiology, Journal Name: Cardiovascular and Interventional Radiology Journal Issue: 3 Vol. 32; ISSN 0174-1551; ISSN CAIRDG
- Country of Publication:
- United States
- Language:
- English
Similar Records
Primary Self-Expandable Nitinol Stent Placement in Focal Lesions of Infrarenal Abdominal Aorta: Long Term Results
Retrieval of a Greenfield IVC Filter Displaced to the Right Brachiocephalic Vein
Endovascular Treatment of Ruptured Abdominal Aortic Aneurysm with Aortocaval Fistula
Journal Article
·
Mon Jan 14 23:00:00 EST 2008
· Cardiovascular and Interventional Radiology
·
OSTI ID:21090673
Retrieval of a Greenfield IVC Filter Displaced to the Right Brachiocephalic Vein
Journal Article
·
Fri Sep 15 00:00:00 EDT 2000
· Cardiovascular and Interventional Radiology
·
OSTI ID:21083585
Endovascular Treatment of Ruptured Abdominal Aortic Aneurysm with Aortocaval Fistula
Journal Article
·
Sun Aug 15 00:00:00 EDT 2010
· Cardiovascular and Interventional Radiology
·
OSTI ID:21428972