Title: A multi-channel monolithic Ge detector system for fluorescence x-ray absorption spectroscopy

Conference ·
OSTI ID:213879
; ; ;  [1]; ; ; ; ;  [2]
  1. Lawrence Berkeley National Lab., CA (United States). Chemical Sciences Div.
  2. Lawrence Berkeley National Lab., CA (United States). Engineering Div.

Construction and performance of a monolithic quad-pixel Ge detector for fluorescence x-ray absorption spectroscopy (XAS) at synchrotron radiation sources are described. The detector semiconductor element has an active surface area of 4.0 cm{sup 2} which is electrically separated into four 1.0 cm{sup 2} pixels, with little interfacial dead volume. Spatial response of the array shows that cross-talk between adjacent pixels is < 10% for 5.9 keV photons that fall within 0.5 mm of the pixel boundaries. The detector electronics system uses pre-amplifiers built at LBNL with commercial Tennelec Model TC 244 amplifiers. Using an {sup 55}Fe test source (MnK{sub {alpha}}, 5.9 keV), energy resolution of better than 200 eV is achieved with a 4 {mu}sec peaking time. At 0.5 {mu}sec peaking time, pulse pileup results in a 75% throughput efficiency for an incoming count rate of 100 kHz. Initial XAS fluoresncece measurements at the beamline 4 wiggler end stations at SSRL show that the detector system has several advantages over commercial x-ray spectrometers for low-concentration counting.

Research Organization:
Lawrence Berkeley National Lab. (LBNL), Berkeley, CA (United States)
Sponsoring Organization:
USDOE, Washington, DC (United States)
DOE Contract Number:
AC03-76SF00098
OSTI ID:
213879
Report Number(s):
LBL-37795; CONF-9510119-34; ON: DE96008561; TRN: 96:010643
Resource Relation:
Conference: SRI `95: synchrotron radiation instrumentation symposium and the 7. users meeting for the advanced photon source (APS), Argonne, IL (United States), 16-20 Oct 1995; Other Information: PBD: Mar 1995
Country of Publication:
United States
Language:
English

Similar Records

A multichannel monolithic Ge detector system for fluorescence x-ray absorption spectroscopy
Journal Article · Sun Sep 01 00:00:00 EDT 1996 · Review of Scientific Instruments · OSTI ID:213879

High energy resolution x-ray spectrometer for high count rate XRF applications
Conference · Sun Aug 01 00:00:00 EDT 1993 · OSTI ID:213879

Tokamak fusion test reactor prototype x-ray pulse-height analyzer diagnostic
Journal Article · Wed May 01 00:00:00 EDT 1985 · Rev. Sci. Instrum.; (United States) · OSTI ID:213879