skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Evaluation of the Effectiveness of the Stereotactic Body Frame in Reducing Respiratory Intrafractional Organ Motion Using the Real-Time Tumor-Tracking Radiotherapy System

Journal Article · · International Journal of Radiation Oncology, Biology and Physics
 [1]; ;  [1]; ; ;  [2]; ; ; ; ;  [3]
  1. Department of Medical Physics, Hokkaido University, Sapporo, Hokkaido (Japan)
  2. Radiology Department, Hokkaido University, Sapporo, Hokkaido (Japan)
  3. Graduate School of Medicine, Hokkaido University, Sapporo, Hokkaido (Japan)

Purpose: To evaluate the effectiveness of the stereotactic body frame (SBF), with or without a diaphragm press or a breathing cycle monitoring device (Abches), in controlling the range of lung tumor motion, by tracking the real-time position of fiducial markers. Methods and Materials: The trajectories of gold markers in the lung were tracked with the real-time tumor-tracking radiotherapy system. The SBF was used for patient immobilization and the diaphragm press and Abches were used to actively control breathing and for self-controlled respiration, respectively. Tracking was performed in five setups, with and without immobilization and respiration control. The results were evaluated using the effective range, which was defined as the range that includes 95% of all the recorded marker positions in each setup. Results: The SBF, with or without a diaphragm press or Abches, did not yield effective ranges of marker motion which were significantly different from setups that did not use these materials. The differences in the effective marker ranges in the upper lobes for all the patient setups were less than 1mm. Larger effective ranges were obtained for the markers in the middle or lower lobes. Conclusion: The effectiveness of controlling respiratory-induced organ motion by using the SBF+diaphragm press or SBF + Abches patient setups were highly dependent on the individual patient reaction to the use of these materials and the location of the markers. They may be considered for lung tumors in the lower lobes, but are not necessary for tumors in the upper lobes.

OSTI ID:
21372318
Journal Information:
International Journal of Radiation Oncology, Biology and Physics, Vol. 77, Issue 2; Other Information: DOI: 10.1016/j.ijrobp.2009.08.060; PII: S0360-3016(09)03036-3; Copyright (c) 2010 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; ISSN 0360-3016
Country of Publication:
United States
Language:
English