An active drop counting device using condenser microphone for superheated emulsion detector
- Nuclear and Atomic Physics Division, Saha Institute of Nuclear Physics, 1/AF Bidhan Nagar, Kolkata 700064 (India)
- Department of Physics, VIT University, Vellore, Tamilnadu 632014 (India)
An active device for superheated emulsion detector is described. A capacitive diaphragm sensor or condenser microphone is used to convert the acoustic pulse of drop nucleation to electrical signal. An active peak detector is included in the circuit to avoid multiple triggering of the counter. The counts are finally recorded by a microprocessor based data acquisition system. Genuine triggers, missed by the sensor, were studied using a simulated clock pulse. The neutron energy spectrum of {sup 252}Cf fission neutron source was measured using the device with R114 as the sensitive liquid and compared with the calculated fission neutron energy spectrum of {sup 252}Cf. Frequency analysis of the detected signals was also carried out.
- OSTI ID:
- 21266710
- Journal Information:
- Review of Scientific Instruments, Vol. 79, Issue 11; Other Information: DOI: 10.1063/1.3020698; (c) 2008 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
- Country of Publication:
- United States