Title: Evaluation of the sensitivity and fading characteristics of an image plate system for x-ray diagnostics

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.3013123· OSTI ID:21266706
; ;  [1]
  1. Plasma Physics Department, AWE Aldermaston, Reading, Berkshire RG7 4PR (United Kingdom)

Image plates (IPs) are a reusable recording media capable of detecting ionizing radiation, used to diagnose x-ray emission from laser-plasma experiments. Due to their superior performance characteristics in x-ray applications [C. C. Bradford, W. W. Peppler, and J. T. Dobbins III, Med. Phys. 26, 27 (1999) and J. Digit. Imaging. 12, 54 (1999)], the Fuji Biological Analysis System (BAS) IPs are fielded on x-ray diagnostics for the HELEN laser by the Plasma Physics Department at AWE. The sensitivities of the Fuji BAS IPs have been absolutely calibrated for absolute measurements of x-ray intensity in the energy range of 0-100 keV. In addition, the Fuji BAS IP fading as a function of time was investigated. We report on the characterization of three Fuji BAS IP responses to x-rays using a radioactive source, and discrete x-ray line energies generated by the Excalibur soft x-ray facility and the Defense Radiological Standards Centre filter-fluorescer hard x-ray system at AWE.

OSTI ID:
21266706
Journal Information:
Review of Scientific Instruments, Vol. 79, Issue 11; Other Information: DOI: 10.1063/1.3013123; (c) 2008 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748