X-ray diagnostic calibration with the tabletop laser facility EQUINOX
- CEA-DIF, Bruyeres-le-Chatel, 91297 Arpajon (France)
The broadband x-ray emission of a target irradiated by a laser can be used to check the calibration of detectors. At CEA-DIF we have a tabletop picosecond laser facility called EQUINOX with 0.3 J at 800 nm. The laser is focused inside a target chamber onto a solid target and produces bright radiation in the 100-2000 eV spectral range. The x-ray source is routinely monitored with a pinhole camera for source dimension measurement and with x-ray diodes for flux measurement. In addition an x-ray transmission grating spectrometer, a crystal spectrometer, and a single count charge coupled device camera measure the x-ray spectrum between 100 eV and 15 keV. The absolute calibration of those sets of spectrometers allows us to fully characterize x-ray emission spectra. Typical duration is less than 100 ps. The spectrum can be tuned by changing target material, pulse length, and x-ray filters. An application to checking the calibration of x-ray diodes used in the broad band spectrometer DMX with single shots will be presented.
- OSTI ID:
- 21266592
- Journal Information:
- Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 10 Vol. 79; ISSN 0034-6748; ISSN RSINAK
- Country of Publication:
- United States
- Language:
- English