Title: Monochromatic x-ray sampling streak imager for fast-ignitor plasma observation

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.2965145· OSTI ID:21266570
; ; ; ; ; ;  [1]
  1. Institute of Laser Engineering, Osaka University, 2-6 Yamada-Oka, Suita, Osaka 565-0871 (Japan)

Ultrafast two-dimensional (2D) x-ray imaging is required to investigate the dynamics of fast-heated core plasma in inertial confinement fusion research. A novel x-ray imager, consisting of two toroidally bent Bragg crystals and an ultrafast 2D x-ray imaging camera, has been demonstrated. Sequential and 2D monochromatic x-ray images of laser-imploded core plasma were obtained with a temporal resolution of 20 ps, a spatial resolution of 31 {mu}m, and a spectral resolution of over 200, simultaneously.

OSTI ID:
21266570
Journal Information:
Review of Scientific Instruments, Vol. 79, Issue 10; Conference: HTPD08: 17. topical conference on high-temperature plasma diagnostics, Albuquerque, NM (United States), 11-15 May 2008; Other Information: DOI: 10.1063/1.2965145; (c) 2008 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English