Dichroic filters to protect milliwatt far-infrared detectors from megawatt ECRH radiation

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.2965777· OSTI ID:21266534
;  [1]; ;  [2]
  1. Institut fuer Energieforschung-Plasmaphysik, Forschungszentrum Juelich, Association EURATOM-FZJ, Trilateral Euregio Cluster, 52425 Juelich (Germany)
  2. I. Physikalisches Institut, Universitaet zu Koeln, Zuelpicher Str. 77, 50937 Koeln (Germany)
Dichroic filters have been used to shield effectively the far infrared (FIR) detectors at the interferometer/polarimeter on TEXTOR. The filters consist of metal foils with regular holes, the hole diameter, the mutual spacing and the thickness of the foils are chosen to transmit radiation at the design frequency with transmission >90%. The attenuation at the low frequency end of the bandpass filter is about 30 dB per octave, the high frequency transmission is between 20% and 40%. The filters have been used to block the stray radiation from the megawatt microwave heating beam to the detectors of the FIR interferometer, operating with power on the detector in the milliwatt range. If required, the low frequency attenuation can be still enhanced, without compromising the transmission in the passband. The FIR interferometer used for plasma density and position control is no longer disturbed by electromagnetic waves used for plasma heating.
OSTI ID:
21266534
Journal Information:
Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 10 Vol. 79; ISSN 0034-6748; ISSN RSINAK
Country of Publication:
United States
Language:
English

Similar Records

Far-infrared dichroic bandpass filters
Journal Article · Wed May 15 00:00:00 EDT 1985 · Appl. Opt.; (United States) · OSTI ID:5769585

Design and performance of the collective Thomson scattering receiver at ASDEX Upgrade
Journal Article · Sat Jan 14 23:00:00 EST 2012 · Review of Scientific Instruments · OSTI ID:22063772

Far-infrared polarimetric characterization of metallic mirrors exposed to a tokamak plasma
Journal Article · Thu Dec 14 23:00:00 EST 2006 · Review of Scientific Instruments · OSTI ID:20951078