Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

SI traceable calibration of an instrumented indentation sensor spring constant using electrostatic force

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.2987695· OSTI ID:21266443
; ; ; ;  [1]
  1. Manufacturing Engineering Laboratory, National Institute of Standards and Technology, Gaithersburg, Maryland 20899 (United States)

We present a measurement scheme for creating reference electrostatic forces that are traceable to the International System of Units. This scheme yields reference forces suitable for calibrating the force sensitivity of instrumented indentation machines and atomic force microscopes. Forces between 10 and 200 {mu}N were created and expressed in terms of the voltage, length, and capacitance between a pair of interacting electrodes. The electrodes comprised an electrically conductive sphere mounted as a tip on an instrumented indentation sensor, and a planar counterelectrode fixed to a sample stage in close proximity to the sphere. For comparison, we applied mechanical forces of similar magnitudes, first using deadweights and then using a reference force sensor. The deflection of the sensor due to the various applied forces was measured using an interferometer. A spring constant for the sensor was computed from the observed records of force versus displacement. Each procedure yielded a relative standard uncertainty of approximately 1%; however, the electrostatic technique is scalable and could provide traceable reference forces as small as a few hundred piconewtons, a range far below anything yet achieved using deadweights.

OSTI ID:
21266443
Journal Information:
Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 9 Vol. 79; ISSN 0034-6748; ISSN RSINAK
Country of Publication:
United States
Language:
English

Similar Records

Traceable Micro-Force Sensor for Instrumented Indentation Calibration
Conference · Mon Apr 02 00:00:00 EDT 2007 · OSTI ID:908085

Electrostatically actuated encased cantilevers
Journal Article · Tue May 08 00:00:00 EDT 2018 · Beilstein Journal of Nanotechnology · OSTI ID:1460304