Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Comparative Analysis of Different Measurement Techniques for MLC Characterization: Preliminary Results

Journal Article · · AIP Conference Proceedings
DOI:https://doi.org/10.1063/1.2979309· OSTI ID:21149298
; ;  [1]; ;  [2]; ;  [3]
  1. Laboratorio de Fisica Medica, Instituto Nacional de Neurologia y Neurocirugia, Insurgentes Sur 3877, Tlalpan, 14269, Mexico D. F. (Mexico)
  2. Instituto de Fisica, UNAM, Ciudad Universitaria, 04510, Mexico, D. F. (Mexico)
  3. Unidad de Radioneurocirugia, Instituto Nacional de Neurologia y Neurocirugia, Insurgentes Sur 3877, Tlalpan, 14269, Mexico D. F. (Mexico)
Radiation transmission, leakage and beam penumbra are essential dosimetric parameters related to the commissioning of a multileaf collimation system. This work shows a comparative analysis of commonly used film detectors: X-OMAT V2 and EDR2 radiographic films, and GafChromic EBT registered radiochromic film. The results show that X-OMAT over-estimates radiation leakage and 80-20% beam penumbra. However, according to the reference values reported by the manufacturer for these dosimetric parameters, all three films are adequate for MLC dosimetric characterization, but special care must be taken when X-OMAT V2 film is used due to its low energy photon dependence.
OSTI ID:
21149298
Journal Information:
AIP Conference Proceedings, Journal Name: AIP Conference Proceedings Journal Issue: 1 Vol. 1032; ISSN APCPCS; ISSN 0094-243X
Country of Publication:
United States
Language:
English

Similar Records

Small Radiation Beam Dosimetry for Radiosurgery of Trigeminal Neuralgia: One Case Analysis
Journal Article · Mon Aug 11 00:00:00 EDT 2008 · AIP Conference Proceedings · OSTI ID:21149278

Measurement of high energy x-ray beam penumbra with Gafchromic trade mark sign EBT radiochromic film
Journal Article · Tue Aug 15 00:00:00 EDT 2006 · Medical Physics · OSTI ID:20853391

SU-F-T-366: Dosimetric Parameters Enhancement of 120-Leaf Millennium MLC Using EGSnrc and IAEA Phase-Space Data
Journal Article · Wed Jun 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22648964