Evaluation of a Single-Isocenter Technique for Axillary Radiotherapy in Breast Cancer
Journal Article
·
· Medical Dosimetry
- Netherlands Cancer Institute-Antoni van Leeuwenhoek Hospital, Department of Radiation Oncology, Amsterdam (Netherlands)
The aim of this study was to develop a technique for axillary radiotherapy that minimizes the risk of radiation-induced damage to the surrounding normal tissue (i.e., arm, shoulder, lung, esophagus, and spinal cord) while keeping the risk of a nodal recurrence to a minimum. A planning study was performed in 20 breast cancer patients. The target volume of the axillary treatment encompassed the periclavicular and axillary lymph node areas. The 3-dimensional (3D) computed tomography (CT) information in this study was used to outline the lymph node areas and the organs at risk (i.e., the esophagus, spinal cord, brachial plexus, and lung). A conventional AP-PA technique (with a transmission plate placed in the AP beam) was evaluated. In addition, a new single-isocenter technique consisting of AP/PA fields using a gantry rotation of {+-}20 deg. and a medial AP segment was developed. Both techniques were compared by evaluation of the calculated dose distributions and the dose-volume histograms of the target volume and surrounding organs at risk. The field borders and humeral shielding were redefined based on the 3D anatomical references. Adapting the humeral shielding reduced the irradiated volume by 19% and might contribute to a reduction of the incidence of arm edema and impairment of shoulder function. The maximum radiation dose in the esophagus and spinal cord was reduced by more than 50% using the single-isocenter technique. The difference between both techniques with respect to the mean doses in the target volume and lung, and the maximum dose in brachial plexus, was not statistically significant. Moreover, the single-isocenter technique allowed a fast and easy treatment preparation and reduced the execution time considerably (with approximately 10 minutes per fraction)
- OSTI ID:
- 21140786
- Journal Information:
- Medical Dosimetry, Journal Name: Medical Dosimetry Journal Issue: 3 Vol. 33; ISSN 0958-3947; ISSN MEDOEJ
- Country of Publication:
- United States
- Language:
- English
Similar Records
An optimized posterior axillary boost technique in radiation therapy to supraclavicular and axillary lymph nodes: A comparative study
Consideration of Dose Limits for Organs at Risk of Thoracic Radiotherapy: Atlas for Lung, Proximal Bronchial Tree, Esophagus, Spinal Cord, Ribs, and Brachial Plexus
Breast Cancer Regional Radiation Fields for Supraclavicular and Axillary Lymph Node Treatment: Is a Posterior Axillary Boost Field Technique Optimal?
Journal Article
·
Mon Dec 31 23:00:00 EST 2012
· Medical Dosimetry
·
OSTI ID:22262843
Consideration of Dose Limits for Organs at Risk of Thoracic Radiotherapy: Atlas for Lung, Proximal Bronchial Tree, Esophagus, Spinal Cord, Ribs, and Brachial Plexus
Journal Article
·
Wed Nov 30 23:00:00 EST 2011
· International Journal of Radiation Oncology, Biology and Physics
·
OSTI ID:22054503
Breast Cancer Regional Radiation Fields for Supraclavicular and Axillary Lymph Node Treatment: Is a Posterior Axillary Boost Field Technique Optimal?
Journal Article
·
Fri May 01 00:00:00 EDT 2009
· International Journal of Radiation Oncology, Biology and Physics
·
OSTI ID:21276775