Image-Guided Radiotherapy for Cervix Cancer: High-Tech External Beam Therapy Versus High-Tech Brachytherapy

Journal Article · · International Journal of Radiation Oncology, Biology and Physics
 [1]; ; ; ;  [1]
  1. Department of Radiotherapy, Medical University Vienna, Vienna (Austria)

Purpose: Many studies comparing external-beam therapy (EBT) and brachytherapy (BT) are biased because advanced EBT is compared with conventional BT. This study compares high-tech EBT against high-tech BT. Methods and Materials: Nine patients were selected with locally advanced cervix cancer, representing typical clinical situations according to initial tumor extension and response after EBT. Patients were treated either with intracavitary, combined interstitial/intracavitary, or complex interstitial BT. Gross tumor volume, high-risk clinical target volume (CTV), intermediate-risk CTV, bladder, rectum, and sigmoid were delineated. Magnetic resonance-guided BT planning was manually optimized with respect to organ dose limits. Margins (3 and 5 mm) were added to BT CTVs to construct planning target volumes (PTVs) for EBT. Inversely planned EBT with photons (IMRT) and protons (IMPT) was challenged to deliver the highest possible doses to PTVs while respecting D{sub 1cc} and D{sub 2cc} limits from BT, assuming the same fractionation (4 x 7 Gy). The D90 for target structures and normal tissue volumes receiving fractionated doses between 3 and 7 Gy were compared. Results: High-risk CTV doses depended on the clinical situation and radiation quality. If IMRT was limited to D{sub 2cc} and D{sub 1cc} from BT, the D90 for high-risk PTV and intermediate-risk PTV was mostly lower. Volumes receiving 60 Gy (in equivalent dose in 20 Gy fractions) were approximately twice as large for IMRT compared with BT. For IMPT, this volume ratio was lower. Planning target volume doses of IMPT plans with 3-mm margins were comparable to those with BT. Gross tumor volume doses were mostly lower for both IMRT and IMPT. Conclusion: For benchmarking high-tech EBT, high-tech BT techniques have to be used. For cervix cancer boost treatments, both IMRT and IMPT seem to be inferior to advanced BT.

OSTI ID:
21124378
Journal Information:
International Journal of Radiation Oncology, Biology and Physics, Journal Name: International Journal of Radiation Oncology, Biology and Physics Journal Issue: 4 Vol. 71; ISSN IOBPD3; ISSN 0360-3016
Country of Publication:
United States
Language:
English

Similar Records

Parametrial Boost Using Midline Shielding Results in an Unpredictable Dose to Tumor and Organs at Risk in Combined External Beam Radiotherapy and Brachytherapy for Locally Advanced Cervical Cancer
Journal Article · Fri Apr 01 00:00:00 EDT 2011 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:21491707

Dosimetric Comparison of High-Dose-Rate Brachytherapy and Intensity-Modulated Radiation Therapy as a Boost to the Prostate
Journal Article · Thu Jan 14 23:00:00 EST 2010 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:21367614

A Dose-Volume Analysis of Magnetic Resonance Imaging-Aided High-Dose-Rate Image-Based Interstitial Brachytherapy for Uterine Cervical Cancer
Journal Article · Thu Jul 01 00:00:00 EDT 2010 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:21436035