Cone Beam CT Image Guidance for Intracranial Stereotactic Treatments: Comparison With a Frame Guided Set-Up
Journal Article
·
· International Journal of Radiation Oncology, Biology and Physics
- Medical Physics Department, Casa di Cura S. Chiara, Firenze (Italy)
- Clinical Radiobiological Institute, University of Firenze, Firenze (Italy)
- Radiotherapy Department, Casa di Cura S. Chiara, Firenze (Italy)
- Medical Physics Department, Vicenza Hospital, Vicenza (Italy)
Purpose: An analysis is performed of the setup errors measured by a kV cone beam computed tomography (CBCT) for intracranial stereotactic radiotherapy (SRT) patients immobilized by a thermoplastic mask and a bite-block and positioned using stereotactic coordinates. We evaluated the overall positioning precision and accuracy of the immobilizing and localizing systems. The potential of image-guided radiotherapy to replace stereotactic methods is discussed. Methods and Materials: Fifty-seven patients received brain SRT. After a frame-guided setup, before each fraction (131 fractions), a CBCT was acquired and the detected displacements corrected online. Translational and rotational errors were analyzed calculating overall mean and standard deviation. A separate analysis was performed for bite-block (in conjunction with mask) and for simple thermoplastic mask. Interobserver variability for CBCT three-dimensional registration was assessed. The residual error after correction and intrafractional motion were calculated. Results: The mean module of the three-dimensional displacement vector was 3.0 {+-} 1.4 mm. Setup errors for bite block and mask were smaller (2.9 {+-} 1.3 mm) than those for thermoplastic mask alone (3.2 {+-} 1.5 mm), but statistical significance was not reached (p = 0.15). Interobserver variability was negligible. The maximum margin calculated for residual errors and intra fraction motion was small but not negligible (1.57 mm). Conclusions: Considering the detected setup errors, daily image guidance is essential for the efficacy of SRT treatments when mask immobilization is used, and even when a bite-block is used in conjunction. The frame setup is still used as a starting point for the opportunity of rotational corrections. Residual margins after on-line corrections must be evaluated.
- OSTI ID:
- 21124334
- Journal Information:
- International Journal of Radiation Oncology, Biology and Physics, Journal Name: International Journal of Radiation Oncology, Biology and Physics Journal Issue: 3 Vol. 71; ISSN IOBPD3; ISSN 0360-3016
- Country of Publication:
- United States
- Language:
- English
Similar Records
Performance of a Novel Repositioning Head Frame for Gamma Knife Perfexion and Image-Guided Linac-Based Intracranial Stereotactic Radiotherapy
Image-Guided Radiotherapy via Daily Online Cone-Beam CT Substantially Reduces Margin Requirements for Stereotactic Lung Radiotherapy
Impact of Immobilization on Intrafraction Motion for Spine Stereotactic Body Radiotherapy Using Cone Beam Computed Tomography
Journal Article
·
Wed Sep 01 00:00:00 EDT 2010
· International Journal of Radiation Oncology, Biology and Physics
·
OSTI ID:21436168
Image-Guided Radiotherapy via Daily Online Cone-Beam CT Substantially Reduces Margin Requirements for Stereotactic Lung Radiotherapy
Journal Article
·
Sat Mar 15 00:00:00 EDT 2008
· International Journal of Radiation Oncology, Biology and Physics
·
OSTI ID:21039824
Impact of Immobilization on Intrafraction Motion for Spine Stereotactic Body Radiotherapy Using Cone Beam Computed Tomography
Journal Article
·
Mon Oct 01 00:00:00 EDT 2012
· International Journal of Radiation Oncology, Biology and Physics
·
OSTI ID:22149540