skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Absolute calibration of photostimulable image plate detectors used as (0.5-20 MeV) high-energy proton detectors

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.2949388· OSTI ID:21124033
; ; ;  [1];  [2]
  1. LULI, Ecole Polytechnique, UMR 7605, CNRS, CEA, UPMC, Route de Saclay, 91128 Palaiseau Cedex (France)
  2. Physics Department, MS 220, University of Nevada, Reno, Nevada 89557 (United States)

In this paper, the absolute calibration of photostimulable image plates (IPs) used as proton detectors is presented. The calibration is performed in a wide range of proton energies (0.5-20 MeV) by exposing simultaneously the IP and calibrated detectors (radiochromic films and solid state detector CR39) to a source of broadband laser-accelerated protons, which are spectrally resolved. The final result is a calibration curve that enables retrieving the proton number from the IP signal.

OSTI ID:
21124033
Journal Information:
Review of Scientific Instruments, Vol. 79, Issue 7; Other Information: DOI: 10.1063/1.2949388; (c) 2008 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English

Similar Records

Absolute response of Fuji imaging plate detectors to picosecond-electron bunches
Journal Article · Fri Jan 15 00:00:00 EST 2010 · Review of Scientific Instruments · OSTI ID:21124033

Computed radiography: Photostimulable phosphor image plate technology
Journal Article · Wed Nov 01 00:00:00 EST 1989 · Radiologic Technology; (USA) · OSTI ID:21124033

Absolute calibration of GafChromic film for very high flux laser driven ion beams
Journal Article · Fri May 03 00:00:00 EDT 2019 · Review of Scientific Instruments · OSTI ID:21124033