Thermoluminescence dosimetry measurements of brachytherapy sources in liquid water
- Radiation Physics, UT M.D. Anderson Cancer Center, 1515 Holcombe Boulevard, Unit 94, Houston, Texas 77030 (United States)
Radiation therapy dose measurements are customarily performed in liquid water. The characterization of brachytherapy sources is, however, generally based on measurements made with thermoluminescence dosimeters (TLDs), for which contact with water may lead to erroneous readings. Consequently, most dosimetry parameters reported in the literature have been based on measurements in water-equivalent plastics, such as Solid Water. These previous reports employed a correction factor to transfer the dose measurements from a plastic phantom to liquid water. The correction factor most often was based on Monte Carlo calculations. The process of measuring in a water-equivalent plastic phantom whose exact composition may be different from published specifications, then correcting the results to a water medium leads to increased uncertainty in the results. A system has been designed to enable measurements with TLDs in liquid water. This system, which includes jigs to support water-tight capsules of lithium fluoride in configurations suitable for measuring several dosimetric parameters, was used to determine the correction factor from water-equivalent plastic to water. Measurements of several {sup 125}I and {sup 131}Cs prostate brachytherapy sources in liquid water and in a Solid Water phantom demonstrated a correction factor of 1.039{+-}0.005 at 1 cm distance. These measurements are in good agreement with a published value of this correction factor for an {sup 125}I source.
- OSTI ID:
- 21120914
- Journal Information:
- Medical Physics, Journal Name: Medical Physics Journal Issue: 9 Vol. 35; ISSN 0094-2405; ISSN MPHYA6
- Country of Publication:
- United States
- Language:
- English
Similar Records
Dosimetric characterization of a {sup 131}Cs brachytherapy source by thermoluminescence dosimetry in liquid water
Comparison of solid phantoms with water for dosimetry of /sup 125/I brachytherapy sources
Determination of the intrinsic energy dependence of LiF:Mg,Ti thermoluminescent dosimeters for {sup 125}I and {sup 103}Pd brachytherapy sources relative to {sup 60}Co
Journal Article
·
Sun Dec 14 23:00:00 EST 2008
· Medical Physics
·
OSTI ID:22095301
Comparison of solid phantoms with water for dosimetry of /sup 125/I brachytherapy sources
Journal Article
·
Thu Sep 01 00:00:00 EDT 1988
· Med. Phys.; (United States)
·
OSTI ID:7063842
Determination of the intrinsic energy dependence of LiF:Mg,Ti thermoluminescent dosimeters for {sup 125}I and {sup 103}Pd brachytherapy sources relative to {sup 60}Co
Journal Article
·
Sun Dec 14 23:00:00 EST 2014
· Medical Physics
·
OSTI ID:22403186