MLC quality assurance using EPID: A fitting technique with subpixel precision

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.2919560· OSTI ID:21120754
; ;  [1]
  1. Department of Radiation Oncology, Washington University School of Medicine, 4921 Parkview Place, St. Louis, Missouri 63110 (United States)
Amorphous silicon based electronic portal imaging devices (EPIDs) have been shown to be a good alternative to radiographic film for routine quality assurance (QA) of multileaf collimator (MLC) positioning accuracy. In this work, we present a method of acquiring an EPID image of a traditional strip-test image using analytical fits of the interleaf and leaf abutment image signatures. After exposure, the EPID image pixel values are divided by an open field image to remove EPID response and radiation field variations. Profiles acquired in the direction orthogonal to the leaf motion exhibit small peaks caused by interleaf leakage. Gaussian profiles are fitted to the interleaf leakage peaks, the results of which are, using multiobjective optimization, used to calculate the image rotational angle with respect to the collimator axis of rotation. The relative angle is used to rotate the image to align the MLC leaf travel to the image pixel axes. The leaf abutments also present peaks that are fitted by heuristic functions, in this case modified Lorentzian functions. The parameters of the Lorentzian functions are used to parameterize the leaf gap width and positions. By imaging a set of MLC fields with varying gaps forming symmetric and asymmetric abutments, calibration curves with regard to relative peak height (RPH) versus nominal gap width are obtained. Based on this calibration data, the individual leaf positions are calculated to compare with the nominal programmed positions. The results demonstrate that the collimator rotation angle can be determined as accurate as 0.01 deg. . A change in MLC gap width of 0.2 mm leads to a change in RPH of about 10%. For asymmetrically produced gaps, a 0.2 mm MLC leaf gap width change causes 0.2 pixel peak position change. Subpixel resolution is obtained by using a parameterized fit of the relatively large abutment peaks. By contrast, for symmetrical gap changes, the peak position remains unchanged with a standard deviation of 0.05 pixels, or 0.026 mm. A trial run of 36 test images, each with gap widths varying from 0.4 to 1.4 mm, were used to analyze 8640 abutments. The leaf position variations were detected with a precision of 0.1 mm at a 95% confidence level, with a mean of 0.04 mm and a standard deviation of 0.03 mm. The proposed method is robust and minimizes the effect of image noise and pixel size and may help physicists to establish reliable and reasonable action levels in routine MLC QA.
OSTI ID:
21120754
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 6 Vol. 35; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

EPID-based verification of the MLC performance for dynamic IMRT and VMAT
Journal Article · Mon Oct 15 00:00:00 EDT 2012 · Medical Physics · OSTI ID:22099044

SU-G-TeP4-03: A Multileaf Collimator Calibration and Quality Assurance Technique Using An Electronic Portal Imaging Device
Journal Article · Wed Jun 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22649466

SU-E-T-646: Quality Assurance of Truebeam Multi-Leaf Collimator Using a MLC QA Phantom
Journal Article · Mon Jun 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:22538155