Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Relationship between isotope half-life and prostatic edema for optimal prostate dose coverage in permanent seed implants

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.2900722· OSTI ID:21120695
; ; ; ;  [1]
  1. Departement de Radio-Oncologie et Centre de Recherche en Cancerologie, Hotel-dieu de Quebec, CHUQ, 11 Cote du Palais, Quebec, Quebec G1R 2J6 (Canada) and Departement de physique, de genie physique et d'optique, Universite Laval, Quebec, Quebec G1K 7P4 (Canada)
The robustness of treatment planning to prostatic edema for three different isotopes ({sup 125}I, {sup 103}Pd, and {sup 131}Cs) is explored using dynamical dose calculations on 25 different clinical prostate cases. The treatment plans were made using the inverse planning by simulated annealing (IPSA) algorithm. The prescription was 144, 127, and 125 Gy for {sup 125}I, {sup 131}Cs, and {sup 103}Pd, respectively. For each isotope, three dose distribution schemes were used to impose different protection levels to the urethra: V{sub 120}=0%, V{sub 150}=0%, and V{sub 150}=30%. Eleven initial edema values were considered ranging from 1.0 (no edema) to 2.0 (100%). The edema was assumed to resolve exponentially with time. The prostate volume, seed positions, and seed activity were dynamically tracked to produce the final dose distribution. Edema decay half-lives of 10, 30, and 50 days were used. A total of 675 dynamical calculations were performed for each initial edema value. For the {sup 125}I isotope, limiting the urethra V{sub 120} to 0% leads to a prostate D{sub 90} under 140 Gy for initial edema values above 1.5. Planning with urethra V{sub 150} at 0% provides a good response to the edema; the prostate D{sub 90} remains higher than 140 Gy for edema values up to 1.8 and a half-life of 30 days or less. For {sup 103}Pd, the prostate D{sub 90} is under 97% of the prescription dose for approximately 66%, 40%, and 30% of edema values for urethra V{sub 120}=0%, V{sub 150}=0%, and V{sub 150}=30%, respectively. Similar behavior is seen for {sup 131}Cs and the center of the prostate becomes 'cold' for almost all edema scenarios. The magnitude of the edema following prostate brachytherapy, as well as the half-life of the isotope used and that of the edema resorption, all have important impacts on the dose distribution. The {sup 125}I isotope with its longer half-life is more robust to prostatic edema. Setting up good planning objectives can provide an adequate compromise between organ doses and robustness. This is even more important since seed misplacements will contribute to further degrade dose coverage.
OSTI ID:
21120695
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 5 Vol. 35; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

Prescription dose in permanent {sup 131}Cs seed prostate implants
Journal Article · Mon Aug 15 00:00:00 EDT 2005 · Medical Physics · OSTI ID:20726263

Determination of the prescription dose for biradionuclide permanent prostate brachytherapy
Journal Article · Sun Dec 14 23:00:00 EST 2008 · Medical Physics · OSTI ID:22095277

Comparison of Dosimetric and Biologic Effective Dose Parameters for Prostate and Urethra Using {sup 131}Cs and {sup 125}I for Prostate Permanent Implant Brachytherapy
Journal Article · Mon Sep 01 00:00:00 EDT 2008 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:21124447