Title: Prostate intrafraction motion evaluation using kV fluoroscopy during treatment delivery: A feasibility and accuracy study

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.2899998· OSTI ID:21120684
;  [1]
  1. Department of Radiation Oncology, Wayne State University, Detroit, Michigan 48201 and Department of Radiation Oncology, William Beaumont Hospital, Royal Oak, Michigan 48073 (United States)

Margin reduction for prostate radiotherapy is limited by uncertainty in prostate localization during treatment. We investigated the feasibility and accuracy of measuring prostate intrafraction motion using kV fluoroscopy performed simultaneously with radiotherapy. Three gold coils used for target localization were implanted into the patient's prostate gland before undergoing hypofractionated online image-guided step-and-shoot intensity modulated radiation therapy (IMRT) on an Elekta Synergy linear accelerator. At each fraction, the patient was aligned using a cone-beam computed tomography (CBCT), after which the IMRT treatment delivery and fluoroscopy were performed simultaneously. In addition, a post-treatment CBCT was acquired with the patient still on the table. To measure the intrafraction motion, we developed an algorithm to register the fluoroscopy images to a reference image derived from the post-treatment CBCT, and we estimated coil motion in three-dimensional (3D) space by combining information from registrations at different gantry angles. We also detected the MV beam turning on and off using MV scatter incident in the same fluoroscopy images, and used this information to synchronize our intrafraction evaluation with the treatment delivery. In addition, we assessed the following: the method to synchronize with treatment delivery, the dose from kV imaging, the accuracy of the localization, and the error propagated into the 3D localization from motion between fluoroscopy acquisitions. With 0.16 mAs/frame and a bowtie filter implemented, the coils could be localized with the gantry at both 0 deg. and 270 deg. with the MV beam off, and at 270 deg. with the MV beam on when multiple fluoroscopy frames were averaged. The localization in two-dimensions for phantom and patient measurements was performed with submillimeter accuracy. After backprojection into 3D the patient localization error was (-0.04{+-}0.30) mm, (0.09{+-}0.36) mm, and (0.03{+-}0.68) mm in the right-left (RL), anterior-posterior (AP), and superior-inferior (SI) axes, respectively. Simulations showed that while oscillating (stationary) motion cannot be effectively represented in 3D, linearly drifting (nonstationary) motion is detectable with good accuracy. These results show that measuring prostate intrafraction motion using a single kV imager during radiotherapy is feasible and can be performed with acceptable accuracy.

OSTI ID:
21120684
Journal Information:
Medical Physics, Vol. 35, Issue 5; Other Information: DOI: 10.1118/1.2899998; (c) 2008 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English

Similar Records

Optimized Hybrid Megavoltage-Kilovoltage Imaging Protocol for Volumetric Prostate Arc Therapy
Journal Article · Fri Oct 01 00:00:00 EDT 2010 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:21120684

Inferences About Prostate Intrafraction Motion From Pre- and Posttreatment Volumetric Imaging
Journal Article · Tue Sep 01 00:00:00 EDT 2009 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:21120684

The first clinical treatment with kilovoltage intrafraction monitoring (KIM): A real-time image guidance method
Journal Article · Thu Jan 15 00:00:00 EST 2015 · Medical Physics · OSTI ID:21120684