The impact of MLC transmitted radiation on EPID dosimetry for dynamic MLC beams
- Royal North Shore Hospital, Sydney (Australia)
The purpose of this study was to experimentally quantify the change in response of an amorphous silicon (a-Si) electronic portal imaging device (EPID) to dynamic multileaf collimator (dMLC) beams with varying MLC-transmitted dose components and incorporate the response into a commercial treatment planning system (TPS) EPID prediction model. A combination of uniform intensity dMLC beams and static beams were designed to quantify the effect of MLC transmission on EPID response at the central axis of 10x10 cm{sup 2} beams, at off-axis positions using wide dMLC beam profiles, and at different field sizes. The EPID response to MLC transmitted radiation was 0.79{+-}0.02 of the response to open beam radiation at the central axis of a 10x10 cm{sup 2} field. The EPID response to MLC transmitted radiation was further reduced relative to the open beam response with off-axis distance. The EPID response was more sensitive to field size changes for MLC transmitted radiation compared to open beam radiation by a factor of up to 1.17 at large field sizes. The results were used to create EPID response correction factors as a function of the fraction of MLC transmitted radiation, off-axis distance, and field size. Software was developed to apply the correction factors to each pixel in the TPS predicted EPID image. The corrected images agreed more closely with the measured EPID images in areas of intensity modulated fields with a large fraction of MLC transmission and, as a result the accuracy of portal dosimetry with a-Si EPIDs can be improved. Further investigation into the detector response function and the radiation source model are required to achieve improvements in accuracy for the general case.
- OSTI ID:
- 21120635
- Journal Information:
- Medical Physics, Journal Name: Medical Physics Journal Issue: 4 Vol. 35; ISSN 0094-2405; ISSN MPHYA6
- Country of Publication:
- United States
- Language:
- English
Similar Records
Experimental investigation of the response of an amorphous silicon EPID to intensity modulated radiotherapy beams
Dose-to-water conversion for the backscatter-shielded EPID: A frame-based method to correct for EPID energy response to MLC transmitted radiation
An EPID response calculation algorithm using spatial beam characteristics of primary, head scattered and MLC transmitted radiation
Journal Article
·
Wed Nov 14 23:00:00 EST 2007
· Medical Physics
·
OSTI ID:21032849
Dose-to-water conversion for the backscatter-shielded EPID: A frame-based method to correct for EPID energy response to MLC transmitted radiation
Journal Article
·
Fri Aug 15 00:00:00 EDT 2014
· Medical Physics
·
OSTI ID:22409571
An EPID response calculation algorithm using spatial beam characteristics of primary, head scattered and MLC transmitted radiation
Journal Article
·
Sun Jun 15 00:00:00 EDT 2008
· Medical Physics
·
OSTI ID:21120740