Title: Evaluation of B/A nonlinear parameter using an acoustic self-calibrated pulse-echo method

Journal Article · · Applied Physics Letters
DOI:https://doi.org/10.1063/1.2937838· OSTI ID:21102046
;  [1]
  1. Universite Francois Rabelais de Tours, INSERM U930/CNRS FRE2448/IC-CUP, ENI Val de Loire, Rue de la Chocolaterie, BP 3410, 41034 Blois Cedex (France)

The objective of this work is to develop an easy-to-build and robust setup for measuring the nonlinearity parameter B/A in fluids using ultrasound. The method is based on the pulse-echo technique, using a single element broadband acoustic transducer, and requires electrical signal measurements. Results obtained in water and denatured alcohol validate the proposed procedure. The choice of a suitable primary wave frequency is discussed with regard to the transducer sensitivity. Further, the influence of the perturbations introduced by the experimental device nonlinearities, and the role of the reflector on the measured second harmonic field amplitude are investigated.

OSTI ID:
21102046
Journal Information:
Applied Physics Letters, Vol. 92, Issue 21; Other Information: DOI: 10.1063/1.2937838; (c) 2008 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0003-6951
Country of Publication:
United States
Language:
English