Title: Use of a Collagen-Based Device for Closure of Low Brachial Artery Punctures

Journal Article · · Cardiovascular and Interventional Radiology
;  [1];  [2];  [3];  [4]
  1. Rabin Medical Center, Beilinson and Glolda Campuses, Petah Tiqwa, Sakler Faculty of Medicine, Tel Aviv University, Department of Surgery (Israel)
  2. Rabin Medical Center, Beilinson and Glolda Campuses, Petah Tiqwa, Sakler Faculty of Medicine, Tel Aviv University, Department of Radiology (Israel)
  3. Hillel Yaffe Medical Center, Department of Radiology (Israel)
  4. Rabin Medical Center, Beilinson and Glolda Campuses, Petah Tiqwa, Sakler Faculty of Medicine, Tel Aviv University, Department of Interventional Radiology (Israel)

Purpose. To report our experience with the Angioseal vascular closure device for hemostasis of distal brachial artery puncture. Methods. Between September 2003 and August 2005, 64 Angioseal vascular closure devices were inserted in 64 patients (40 men, 24 women; mean age 65 years) immediately after diagnostic or therapeutic arterial angiographies performed through a 5 Fr to 7 Fr sheath via the distal brachial artery. Ultrasound examination of the brachial artery preceded the angiography in all cases and only arteries wider than 4 mm were closed by the Angioseal. In cases of a sonographically evident thin subcutaneous space of the cubital fossa, tissue tumescence, using 1% Lidocaine, was performed prior to the arterial closure. Results. The deployment success rate was 100%. No major complications were encountered; only 2 patients developed puncture site hematoma, and these were followed conservatively. Conclusions. Closure of low brachial artery punctures with the Angioseal is simple and safe. No additional manual compression is required. We recommend its use after brachial artery access interventions, through appropriately wide arteries, to improve early patient ambulation and potentially reduce possible puncture site complications.

OSTI ID:
21091006
Journal Information:
Cardiovascular and Interventional Radiology, Vol. 30, Issue 2; Other Information: DOI: 10.1007/s00270-006-0061-4; Copyright (c) 2007 Springer Science+Business Media, Inc.; www.springer-ny.com; Country of input: International Atomic Energy Agency (IAEA); ISSN 0174-1551
Country of Publication:
United States
Language:
English

Similar Records

A Prospective Randomized Trial Comparing the Safety and Efficacy of Three Commercially Available Closure Devices (Angioseal, Vasoseal and Duett)
Journal Article · Tue Oct 15 00:00:00 EDT 2002 · Cardiovascular and Interventional Radiology · OSTI ID:21091006

Efficacy and Safety of the Angioseal Vascular Closure Device Post Antegrade Puncture
Journal Article · Thu May 15 00:00:00 EDT 2008 · Cardiovascular and Interventional Radiology · OSTI ID:21091006

Immediate Sealing of Arterial Puncture Site Following Femoropopliteal Angioplasty: A Prospective Randomized Trial
Journal Article · Fri Nov 15 00:00:00 EST 1996 · Cardiovascular and Interventional Radiology · OSTI ID:21091006