Polyurethane-Covered Self-expandable Nitinol Stent for Malignant Biliary Obstruction: Preliminary Results

Journal Article · · Cardiovascular and Interventional Radiology
;  [1];  [2]; ;  [1]
  1. Department of Radiology, Chonbuk National University Medical School, 634-18, Keumam Dong, Chonju City, 560-182 (Korea, Republic of)
  2. Department of Internal Medicine, Chonbuk National University Medical School, 634-18, Keumam Dong, Chonju City, 560-182 (Korea, Republic of)
Purpose: To evaluate the immediate and long-term results in ten patients with malignant biliary obstruction using a polyurethane-covered, self-expandable nitinol stent.Methods: A nitinol stent, fully covered with high-elasticity polyurethane, with an unconstrained diameter of 10 mm and a total length of 60-80 mm, was placed transhepatically under fluoroscopic guidance in ten patients. The length of the biliary obstruction varied between 30-50 mm. The follow-up examination included a clinical assessment, serum bilirubin measurement, and ultrasound examinations at one-month intervals in all patients. Results: Initial stent deployment was successful in eight patients. Two patients experienced proximal migration,which was solved by insertion of two, uncovered Wallstents. One patient had the stent removed after two weeks because of severe nausea and vomiting. One patient had delayed migration of the covered stent after 40 weeks. The follow-up duration was 3-40 weeks (Mean: 16.9 weeks). Seven patients died(3-26 weeks) and three patients survived (24-40 weeks). The bilirubin measurements in all patients had significantly decreased one week after stent insertion. The levels of amylase and lipase had did not increase after stent insertion. Conclusions: Apolyurethane-covered, self-expandable nitinol stent can be used effectively and safely in the treatment of malignant biliary obstruction. The preliminary results are favorable, but there is a need for further large studies to determine both long-term survival and patency.
OSTI ID:
21083415
Journal Information:
Cardiovascular and Interventional Radiology, Journal Name: Cardiovascular and Interventional Radiology Journal Issue: 5 Vol. 25; ISSN 0174-1551; ISSN CAIRDG
Country of Publication:
United States
Language:
English

Similar Records

Covered Self-Expanding Transhepatic Biliary Stents:Clinical Pilot Study
Journal Article · Thu Nov 14 23:00:00 EST 1996 · Cardiovascular and Interventional Radiology · OSTI ID:21080561

Percutaneous cholangioscopy in obstructed biliary metal stents
Journal Article · Thu May 15 00:00:00 EDT 1997 · Cardiovascular and Interventional Radiology · OSTI ID:21080218

Polyurethane-Covered Nitinol Strecker Stents as Primary Palliative Treatment of Malignant Biliary Obstruction
Journal Article · Tue Mar 14 23:00:00 EST 2000 · Cardiovascular and Interventional Radiology · OSTI ID:21083669