Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Local Intraarterial Thrombolysis: In Vitro Comparison of Various Infusion Catheters

Journal Article · · Cardiovascular and Interventional Radiology
DOI:https://doi.org/10.1007/S002709900170· OSTI ID:21080470
; ; ; ;  [1];  [2];  [1]
  1. Medizinisches Zentrum fuer Radiologie, Abteilung fuer Strahlendiagnostik, Klinikum der Philipps-Universitaet, D-35043 Marburg (Germany)
  2. Division of Vascular and Interventional Radiology, Georgetown University Hospital, 3800 Reservoir Road, N.W., Washington, DC 20007 (United States)
Purpose: Catheters are compared in vitro to evaluate the efficacy of thrombolysis during urokinase infusion within the thrombus. Methods: Six catheters were introduced individually into human thrombus within a stenotic flow model. Urokinase was infused continuously into the thrombus. To quantify the efficacy of thrombolysis, pressure gradients were recorded proximal and distal to the thrombus and during the course of infusion. Uniformity of lysis was assessed radiographically. Results: The fastest and most homogeneous thrombolysis was achieved with the EDM and the straight-flush catheter, shown by decreasing transthrombotic pressure gradients. All other catheter designs showed less homogeneous and delayed thrombolysis (p{<=} 0.001, Friedmann-Test, Schaich-Hamerle). There was no significant difference in the efficacy of thrombus dissolution between the EDM and the straight-flush catheter (Wilcoxon, matched pairs, p> 0.7). Conclusion: The EDM catheter and the straight flush catheter achieved the most homogeneous and fastest thrombolysis, apparently due to the best urokinase distribution within the thrombus.
OSTI ID:
21080470
Journal Information:
Cardiovascular and Interventional Radiology, Journal Name: Cardiovascular and Interventional Radiology Journal Issue: 5 Vol. 20; ISSN 0174-1551; ISSN CAIRDG
Country of Publication:
United States
Language:
English

Similar Records

Local Intraarterial Thrombolysis: In Vitro Comparison Between Automatic and Manual Pulse-Spray Infusion
Journal Article · Thu Nov 14 23:00:00 EST 1996 · Cardiovascular and Interventional Radiology · OSTI ID:21080483

Initial Results of Catheter-Directed Ultrasound-Accelerated Thrombolysis for Thromboembolic Obstructions of the Aortofemoral Arteries: A Feasibility Study
Journal Article · Sun Apr 15 00:00:00 EDT 2012 · Cardiovascular and Interventional Radiology · OSTI ID:21608566

Catheter-Directed Thrombolysis with Percutaneous Rheolytic Thrombectomy Versus Thrombolysis Alone in Upper and Lower Extremity Deep Vein Thrombosis
Journal Article · Thu Dec 14 23:00:00 EST 2006 · Cardiovascular and Interventional Radiology · OSTI ID:21091088