Purpose: We investigated the relationship between the amount of radiation exposure to the operator during table-side manual-injection angiographic procedures including digital subtraction angiography (DSA) and the operator's position, as well as a simple means to decrease radiation exposure. Methods: Measurement of radiation exposure was carried out with thermoluminescent dosimeters (TLDs) in nine abdominal angiographies. In the first study, radiation exposure during DSA or during fluoroscopy was measured using TLDs placed near the angiographic table. In the second study, radiation exposure to the interventional radiologist was measured during manual-injection DSA at a near and a far operator position. Results: Radiation exposure to the operator received during manual-injection DSA accounted for more than 90% of the total procedural exposure. The exposure to the operator markedly decreased at the far position compared with that at the near position when performing DSA. Conclusion: Manual-injection DSA is the largest contributor to radiation exposure received by the interventional radiologist, therefore, the use of a power injector is always recommended when performing DSA. When manual-injection DSA is necessary, radiologists should position themselves as far away from the patient as possible.

OSTI ID:
21080406
Journal Information:
Cardiovascular and Interventional Radiology, Vol. 21, Issue 3; Other Information: DOI: 10.1007/s002709900252; Copyright (c) 1996 Springer-Verlag New York Inc; Country of input: International Atomic Energy Agency (IAEA); ISSN 0174-1551
Country of Publication:
United States
Language:
English

Similar Records

WE-DE-207A-03: Recent Advances in Devices Used in Neuro--Interventions
Journal Article · Wed Jun 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:21080406

WE-DE-207A-04: Advances in Radiological Neuro-Endovascular Interventional Imaging
Journal Article · Wed Jun 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:21080406

Radiologist and angiographic procedures. Absorbed radiation dose
Journal Article · Tue Sep 01 00:00:00 EDT 1987 · Invest. Radiol.; (United States) · OSTI ID:21080406