skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Elastic Deformation Properties of Implanted Endobronchial Wire Stents in Benign and Malignant Bronchial Disease: A Radiographic In Vivo Evaluation

Journal Article · · Cardiovascular and Interventional Radiology
DOI:https://doi.org/10.1007/S002709900343· OSTI ID:21080332
 [1];  [2];  [1];  [2]
  1. Medizinische Klinik, Klinikum Innenstadt, Ludwig-Maximilians-University, Ziemssentrasse 1, D-80336 Munich (Germany)
  2. Radiologische Klinik, Klinikum Innenstadt, Ludwig-Maximilians-University, Ziemssentrasse 1, D-80336 Munich (Germany)

Purpose: To evaluate the long-term mechanical behavior in vivo of expandable endobronchial wire stents, we imaged three different prostheses in the treatment of tracheobronchial disease. Methods: Six patients with bronchial stenoses (three benign, three malignant) underwent insertion of metallic stents. Two self-expandable Wallstents, two balloon-expandable tantalum Strecker stents and two self-expandable nitinol Accuflex stents were used. Measurements of deformation properties were performed during voluntary cough by means of fluoroscopy, at 1 month and 7-10 months after implantation. The procedures were videotaped, their images digitized and the narrowing of stent diameters calculated at intervals of 20 msec. Results: After stent implantation all patients improved with respect to ventilatory function. Radial stent narrowing during cough reached 53% (Wallstent), 59% (tantalum Strecker stent), and 52% (nitinol Accuflex stent) of the relaxed post-implantation diameter. Stent compression was more marked in benign compared with malignant stenoses. In the long term permanent deformation occurred with the tantalum Strecker stents; the other stents were unchanged. Conclusion: Endobronchial wire stents can be helpful in the treatment of major airway collapse and obstructing bronchial lesions. However, evidence of material fatigue as a possible effect of exposure to recurrent mechanical stress on the flexible mesh tube may limit their long-term use. This seems to be predominantly important in benign bronchial collapse.

OSTI ID:
21080332
Journal Information:
Cardiovascular and Interventional Radiology, Vol. 22, Issue 2; Other Information: DOI: 10.1007/s002709900343; Copyright (c) 1999 Springer-Verlag New York Inc; Country of input: International Atomic Energy Agency (IAEA); ISSN 0174-1551
Country of Publication:
United States
Language:
English