Purpose: To evaluate the long-term mechanical behavior in vivo of expandable endobronchial wire stents, we imaged three different prostheses in the treatment of tracheobronchial disease. Methods: Six patients with bronchial stenoses (three benign, three malignant) underwent insertion of metallic stents. Two self-expandable Wallstents, two balloon-expandable tantalum Strecker stents and two self-expandable nitinol Accuflex stents were used. Measurements of deformation properties were performed during voluntary cough by means of fluoroscopy, at 1 month and 7-10 months after implantation. The procedures were videotaped, their images digitized and the narrowing of stent diameters calculated at intervals of 20 msec. Results: After stent implantation all patients improved with respect to ventilatory function. Radial stent narrowing during cough reached 53% (Wallstent), 59% (tantalum Strecker stent), and 52% (nitinol Accuflex stent) of the relaxed post-implantation diameter. Stent compression was more marked in benign compared with malignant stenoses. In the long term permanent deformation occurred with the tantalum Strecker stents; the other stents were unchanged. Conclusion: Endobronchial wire stents can be helpful in the treatment of major airway collapse and obstructing bronchial lesions. However, evidence of material fatigue as a possible effect of exposure to recurrent mechanical stress on the flexible mesh tube may limit their long-term use. This seems to be predominantly important in benign bronchial collapse.
OSTI ID:
21080332
Journal Information:
Cardiovascular and Interventional Radiology, Journal Name: Cardiovascular and Interventional Radiology Journal Issue: 2 Vol. 22; ISSN 0174-1551; ISSN CAIRDG
Country of Publication:
United States
Language:
English

Similar Records

Treatment of Benign and Malignant Tracheobronchial Obstruction with Metal Wire Stents: Experience with a Balloon-Expandable and a Self-Expandable Stent Type
Journal Article · Sun Aug 15 00:00:00 EDT 2004 · Cardiovascular and Interventional Radiology · OSTI ID:21088255

Metallic Stents for Tracheobronchial Pathology Treatment
Journal Article · Sat Dec 14 23:00:00 EST 2013 · Cardiovascular and Interventional Radiology · OSTI ID:22207970

Treatment of Inoperable Tracheobronchial Obstructive Lesions with the Palmaz Stent
Journal Article · Sun Mar 14 23:00:00 EST 1999 · Cardiovascular and Interventional Radiology · OSTI ID:21080328