We have embarked upon a systematic program to optimize all aspects of small-angle-x-ray-scattering x-ray photon-correlation spectroscopy experiments performed at beamline 8-ID-I at the Advanced Photon Source (APS). We describe two recent efforts, namely better synchronized CCD shutter control and a re-engineered experiment station layout that has resulted in a 70% increase in optical contrast and a greater than 300% increase in experiment stability. Together these two improvements extend the dynamic range of time scales that can be probed at APS beamline 8-ID-I by one decade.

OSTI ID:
21052665
Journal Information:
AIP Conference Proceedings, Vol. 879, Issue 1; Conference: 9. international conference on synchrotron radiation instrumentation, Daegu (Korea, Republic of), 28 May - 2 Jun 2006; Other Information: DOI: 10.1063/1.2436206; (c) 2007 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-243X