Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Improvement of the cine-CT based 4D-CT imaging

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.2794225· OSTI ID:21032859
; ;  [1]
  1. Department of Imaging Physics, M. D. Anderson Cancer Center, University of Texas, Houston, Texas 77030 (United States)
An improved 4D-CT utility has been developed on the GE LightSpeed multislice CT (MSCT) and Discovery PET/CT scanners, which have the cine CT scan capability. Two new features have been added in this 4D-CT over the commercial Advantage 4D-CT from GE. One feature was a new tool for disabling parts of the respiratory signal with irregular respiration and improving the accuracy of phase determination for the respiratory signal from the Varian real-time positioning and monitoring (RPM) system before sorting of the cine CT images into the 4D-CT images. The second feature was to allow generation of the maximum-intensity-projection (MIP), average (AVG) and minimum-intensity-projection (mip) CT images from the cine CT images without a respiratory signal. The implementation enables the assessment of tumor motion in treatment planning with the MIP, AVG, and mip CT images on the GE MSCT and PET/CT scanners without the RPM and the Advantage 4D-CT with a GE Advantage windows workstation. Several clinical examples are included to illustrate this new application.
OSTI ID:
21032859
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 11 Vol. 34; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

TU-F-CAMPUS-J-03: Evaluation of a New GE Device-Less Cine 4D-CT
Journal Article · Mon Jun 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:22570062

Retrospective Analysis of Artifacts in Four-Dimensional CT Images of 50 Abdominal and Thoracic Radiotherapy Patients
Journal Article · Fri Nov 14 23:00:00 EST 2008 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:21172482

SU-C-9A-06: The Impact of CT Image Used for Attenuation Correction in 4D-PET
Journal Article · Sun Jun 01 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22334131