Design and evaluation of a variable aperture collimator for conformal radiotherapy of small animals using a microCT scanner

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.2789498· OSTI ID:21032847
; ; ; ; ; ;  [1]
  1. Department of Radiology Oncology, Molecular Imaging Program at Stanford, Stanford University, Stanford, California 94305 (United States)

Treatment of small animals with radiation has in general been limited to planar fields shaped with lead blocks, complicating spatial localization of dose and treatment of deep-seated targets. In order to advance laboratory radiotherapy toward what is accomplished in the clinic, we have constructed a variable aperture collimator for use in shaping the beam of microCT scanner. This unit can image small animal subjects at high resolution, and is capable of delivering therapeutic doses in reasonable exposure times. The proposed collimator consists of two stages, each containing six trapezoidal brass blocks that move along a frame in a manner similar to a camera iris producing a hexagonal aperture of variable size. The two stages are offset by 30 deg. and adjusted for the divergence of the x-ray beam so as to produce a dodecagonal profile at isocenter. Slotted rotating driving plates are used to apply force to pins in the collimator blocks and effect collimator motion. This device has been investigated through both simulation and measurement. The collimator aperture size varied from 0 to 8.5 cm as the driving plate angle increased from 0 to 41 deg. . The torque required to adjust the collimator varied from 0.5 to 5 N{center_dot}m, increasing with increasing driving plate angle. The transmission profiles produced by the scanner at isocenter exhibited a penumbra of approximately 10% of the collimator aperture width. Misalignment between the collimator assembly and the x-ray source could be identified on the transmission images and corrected by adjustment of the collimator location. This variable aperture collimator technology is therefore a feasible and flexible solution for adjustable shaping of radiation beams for use in small animal radiotherapy as well as other applications in which beam shaping is desired.

OSTI ID:
21032847
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 11 Vol. 34; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

Development of a Micro-Computed Tomography-Based Image-Guided Conformal Radiotherapy System for Small Animals
Journal Article · Wed Sep 01 00:00:00 EDT 2010 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:21436167

Modality comparison for small animal radiotherapy: A simulation study
Journal Article · Tue Jan 14 23:00:00 EST 2014 · Medical Physics · OSTI ID:22250753

Modality comparison for small animal radiotherapy: A simulation study
Journal Article · Tue Jan 14 23:00:00 EST 2014 · Medical Physics · OSTI ID:22250952