A method for calibrating the stiffness of atomic force microscope (AFM) cantilevers is demonstrated using an array of uniform microfabricated reference cantilevers. A series of force-displacement curves was obtained using a commercial AFM test cantilever on the reference cantilever array, and the data were analyzed using an implied Euler-Bernoulli model to extract the test cantilever spring constant from linear regression fitting. The method offers a factor of 5 improvement over the precision of the usual reference cantilever calibration method and, when combined with the Systeme International traceability potential of the cantilever array, can provide very accurate spring constant calibrations.

OSTI ID:
21024152
Journal Information:
Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 8 Vol. 78; ISSN 0034-6748; ISSN RSINAK
Country of Publication:
United States
Language:
English

Similar Records

Accurate spring constant calibration for very stiff atomic force microscopy cantilevers
Journal Article · Thu Nov 14 23:00:00 EST 2013 · Review of Scientific Instruments · OSTI ID:22220328

Spring constant calibration of atomic force microscopy cantilevers with a piezosensor transfer standard
Journal Article · Sat Sep 15 00:00:00 EDT 2007 · Review of Scientific Instruments · OSTI ID:21024285

Precision and accuracy of thermal calibration of atomic force microscopy cantilevers
Journal Article · Tue Aug 15 00:00:00 EDT 2006 · Review of Scientific Instruments · OSTI ID:20853509