Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Investigation of Linac-Based Image-Guided Hypofractionated Prostate Radiotherapy

Journal Article · · Medical Dosimetry
 [1];  [1];  [1];  [1];  [1];  [1];  [1];  [1]
  1. Department of Radiation Oncology, Stanford University School of Medicine, Stanford, CA (United States)
A hypofractionation treatment protocol for prostate cancer was initiated in our department in December 2003. The treatment regimen consists of a total dose of 36.25 Gy delivered at 7.25 Gy per fraction over 10 days. We discuss the rationale for such a prostate hypofractionation protocol and the need for frequent prostate imaging during treatment. The CyberKnife (Accuray Inc., Sunnyvale, CA), a linear accelerator mounted on a robotic arm, is currently being used as the radiation delivery device for this protocol, due to its incorporation of near real-time kV imaging of the prostate via 3 gold fiducial seeds. Recently introduced conventional linac kV imaging with intensity modulated planning and delivery may add a new option for these hypofractionated treatments. The purpose of this work is to investigate the use of intensity modulated radiotherapy (IMRT) and the Varian Trilogy Accelerator with on-board kV imaging (Varian Medical Systems Inc., Palo Alto, CA) for treatment of our hypofractionated prostate patients. The dose-volume histograms and dose statistics of 2 patients previously treated on the CyberKnife were compared to 7-field IMRT plans. A process of acquiring images to observe intrafraction prostate motion was achieved in an average time of about 1 minute and 40 seconds, and IMRT beam delivery takes about 40 seconds per field. A complete 7-field IMRT plan can therefore be imaged and delivered in 10 to 17 minutes. The Varian Trilogy Accelerator with on-board imaging and IMRT is well suited for image-guided hypofractionated prostate treatments. During this study, we have also uncovered opportunities for improvement of the on-board imaging hardware/software implementation that would further enhance performance in this regard.
OSTI ID:
20975203
Journal Information:
Medical Dosimetry, Journal Name: Medical Dosimetry Journal Issue: 2 Vol. 32; ISSN 0958-3947; ISSN MEDOEJ
Country of Publication:
United States
Language:
English

Similar Records

Image-Guided Stereotactic Spine Radiosurgery on a Conventional Linear Accelerator
Journal Article · Thu Apr 01 00:00:00 EDT 2010 · Medical Dosimetry · OSTI ID:21333970

Effects of residual target motion for image-tracked spine radiosurgery
Journal Article · Wed Nov 14 23:00:00 EST 2007 · Medical Physics · OSTI ID:21032858

SU-F-T-601: Equivalence of CyberKnife and TrueBeam STx for Spine Irradiation
Journal Article · Wed Jun 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22649171