Noncontact measurement of electrostatic fields: Verification of modeled potentials within ion mobility spectrometer drift tube designs
- Idaho National Laboratory (INL), Idaho Falls, Idaho 83415-0001 (United States)
The heart of an ion mobility spectrometer is the drift region where ion separation occurs. While the electrostatic potentials within a drift tube design can be modeled, no method for independently validating the electrostatic field has previously been reported. Two basic drift tube designs were modeled using SIMION 7.0 to reveal the expected electrostatic fields: (1) A traditional alternating set of electrodes and insulators and (2) a truly linear drift tube. One version of the alternating electrode/insulator drift tube and two versions of linear drift tubes were then fabricated. The stacked alternating electrodes/insulators were connected through a resistor network to generate the electrostatic gradient in the drift tube. The two linear drift tube designs consisted of two types of resistive drift tubes with one tube consisting of a resistive coating within an insulating tube and the other tube composed of resistive ferrites. The electrostatic fields within each type of drift tube were then evaluated by a noncontact method using a Kelvin-Zisman type electrostatic voltmeter and probe (results for alternative measurement methods provided in supplementary material). The experimental results were then compared with the electrostatic fields predicted by SIMION. Both the modeling and experimental measurements reveal that the electrostatic fields within a stacked ion mobility spectrometer drift tube are only pseudo-linear, while the electrostatic fields within a resistive drift tube approach perfect linearity.
- OSTI ID:
- 20953402
- Journal Information:
- Review of Scientific Instruments, Vol. 78, Issue 3; Other Information: DOI: 10.1063/1.2713424; (c) 2007 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
- Country of Publication:
- United States
- Language:
- English
Similar Records
Comparison of Ion Trajectories in Vacuum and Viscous Environments using SIMION: Insights for Instrument Design
SimELIT: A Novel GUI-Based Comprehensive Ion Trajectory Simulation Software for Mass Spectrometry