Gamma-ray detection efficiency of the microchannel plate installed as an ion detector in the low energy particle instrument onboard the GEOTAIL satellite
Journal Article
·
· Review of Scientific Instruments
- Department of Earth and Planetary Science, University of Tokyo, 7-3-1 Hongo, Bunkyo-ku, Tokyo 113-0033 (Japan)
A microchannel plate (MCP) assembly has been used as an ion detector in the low energy particle (LEP) instrument onboard the magnetospheric satellite GEOTAIL. Recently the MCP assembly has detected gamma rays emitted from an astronomical object and has been shown to provide unique information of gamma rays if they are intense enough. However, the detection efficiency for gamma rays was not measured before launch, and therefore we could not analyze the LEP data quantitatively. In this article, we report the gamma-ray detection efficiency of the MCP assembly. The measured efficiencies are 1.29%{+-}0.71% and 0.21%{+-}0.14% for normal incidence 60 and 662 keV gamma rays, respectively. The incident angle dependence is also presented. Our calibration is crucial to study high energy astrophysical phenomena by using the LEP.
- OSTI ID:
- 20953396
- Journal Information:
- Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 3 Vol. 78; ISSN 0034-6748; ISSN RSINAK
- Country of Publication:
- United States
- Language:
- English
Similar Records
Gamma-radiation registration efficiency of herringbone assembly of microchannel plates
Comparative Response of Microchannel Plate and Channel Electron Multiplier Detectors to Penetrating Radiation in Space
Performance of Au transmission photocathode on a microchannel plate detector
Journal Article
·
Sat Oct 31 23:00:00 EST 1987
· Instrum. Exp. Tech. (Engl. Transl.); (United States)
·
OSTI ID:5179269
Comparative Response of Microchannel Plate and Channel Electron Multiplier Detectors to Penetrating Radiation in Space
Journal Article
·
Thu Oct 01 20:00:00 EDT 2015
· IEEE Transactions on Nuclear Science
·
OSTI ID:1329562
Performance of Au transmission photocathode on a microchannel plate detector
Journal Article
·
Wed Oct 15 00:00:00 EDT 2008
· Review of Scientific Instruments
·
OSTI ID:21266574