Multichannel detectors for profile measurements in clinical proton fields
- Indiana University Cyclotron Facility, 2401 Milo B. Sampson Lane, Bloomington, Indiana 47408 (United States)
Two beam profile measurement detectors have been developed at Indiana University Cyclotron Facility to address the need for a tool to efficiently verify dose distributions created with active methods of clinical proton beam delivery. The multipad ionization chamber (MPIC) has 128 ionization chambers arranged in one plane and is designed to measure lateral profiles in fields up to 38 cm in diameter. The MPIC pads have a 5 mm pitch for fields up to 20 cm in diameter and a 7 mm pitch for larger fields, providing the accuracy of field size determination about 0.5 mm. The multilayer ionization chamber (MLIC) detector contains 122 small-volume ionization chambers stacked at a 1.82 mm step (water-equivalent) for depth-dose profile measurements. The MLIC detector can measure profiles up to 20 cm in depth, and determine the 80% distal dose fall-off with about 0.1 mm precision. Both detectors can be connected to the same set of electronics modules, which comprise the detectors' data acquisition system. The detectors have been tested in clinical proton fields produced with active methods of beam delivery such as uniform scanning and energy stacking. This article describes detector performance tests and discusses their results. The test results indicate that the MPIC and MLIC detectors can be used for dosimetric characterization of clinical proton fields. The detectors offer significant time savings during measurements in actively delivered beams compared with traditional measurements using a water phantom.
- OSTI ID:
- 20953318
- Journal Information:
- Medical Physics, Journal Name: Medical Physics Journal Issue: 7 Vol. 34; ISSN 0094-2405; ISSN MPHYA6
- Country of Publication:
- United States
- Language:
- English
Similar Records
Dual ring multilayer ionization chamber and theory-based correction technique for scanning proton therapy
SU-E-T-561: Development of Depth Dose Measurement Technique Using the Multilayer Ionization Chamber for Spot Scanning Method
Clinical characterization of a proton beam continuous uniform scanning system with dose layer stacking
Journal Article
·
Fri Jul 15 00:00:00 EDT 2016
· Medical Physics
·
OSTI ID:22689468
SU-E-T-561: Development of Depth Dose Measurement Technique Using the Multilayer Ionization Chamber for Spot Scanning Method
Journal Article
·
Sun Jun 01 00:00:00 EDT 2014
· Medical Physics
·
OSTI ID:22369681
Clinical characterization of a proton beam continuous uniform scanning system with dose layer stacking
Journal Article
·
Fri Nov 14 23:00:00 EST 2008
· Medical Physics
·
OSTI ID:22095247