Comparison of Kodak EDR2 and Gafchromic EBT film for intensity-modulated radiation therapy dose distribution verification
Journal Article
·
· Medical Dosimetry
- Department of Radiation Oncology, Apollo Specialty Hospital, Chennai (India) and Department of Physics, Anna University, Chennai (India)
- Department of Radiation Oncology, University of Nebraska Medical Center, Omaha, NE (United States)
- Department of Radiation Oncology, Apollo Specialty Hospital, Chennai (India)
- Department of Physics, Anna University, Chennai (India)
The quantitative dose validation of intensity-modulated radiation therapy (IMRT) plans require 2-dimensional (2D) high-resolution dosimetry systems with uniform response over its sensitive region. The present work deals with clinical use of commercially available self-developing Radio Chromic Film, Gafchromic EBT film, for IMRT dose verification. Dose response curves were generated for the films using a VXR-16 film scanner. The results obtained with EBT films were compared with the results of Kodak extended dose range 2 (EDR2) films. The EBT film had a linear response between the dose range of 0 to 600 cGy. The dose-related characteristics of the EBT film, such as post irradiation color growth with time, film uniformity, and effect of scanning orientation, were studied. There was up to 8.6% increase in the color density between 2 to 40 hours after irradiation. There was a considerable variation, up to 8.5%, in the film uniformity over its sensitive region. The quantitative differences between calculated and measured dose distributions were analyzed using DTA and Gamma index with the tolerance of 3% dose difference and 3-mm distance agreement. The EDR2 films showed consistent results with the calculated dose distributions, whereas the results obtained using EBT were inconsistent. The variation in the film uniformity limits the use of EBT film for conventional large-field IMRT verification. For IMRT of smaller field sizes (4.5 x 4.5 cm), the results obtained with EBT were comparable with results of EDR2 films.
- OSTI ID:
- 20858083
- Journal Information:
- Medical Dosimetry, Journal Name: Medical Dosimetry Journal Issue: 4 Vol. 31; ISSN 0958-3947; ISSN MEDOEJ
- Country of Publication:
- United States
- Language:
- English
Similar Records
Evaluation of GafChromic EBT prototype B for external beam dose verification
Initial investigation on the use of MR spectroscopy and micro-MRI of GAFCHROMIC EBT radiotherapy film
Comparison of the Epson Expression 1680 flatbed and the Vidar VXR-16 Dosimetry PRO trade mark sign film scanners for use in IMRT dosimetry using Gafchromic and radiographic film
Journal Article
·
Mon May 15 00:00:00 EDT 2006
· Medical Physics
·
OSTI ID:20775203
Initial investigation on the use of MR spectroscopy and micro-MRI of GAFCHROMIC EBT radiotherapy film
Journal Article
·
Sat Nov 14 23:00:00 EST 2009
· Medical Physics
·
OSTI ID:22102162
Comparison of the Epson Expression 1680 flatbed and the Vidar VXR-16 Dosimetry PRO trade mark sign film scanners for use in IMRT dosimetry using Gafchromic and radiographic film
Journal Article
·
Sun Jan 14 23:00:00 EST 2007
· Medical Physics
·
OSTI ID:20853931