Intensity modulated radiation therapy with irregular multileaf collimated field: A dosimetric study on the penumbra region with different leaf stepping patterns

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.2388571· OSTI ID:20853829
; ;  [1]
  1. Department of Radiation Oncology, University of Toronto and Radiation Medicine Program, Princess Margaret Hospital, University Health Network, 610 University Avenue, Toronto, ON N2G 1G3 (Canada) and Department of Physics, University of Waterloo, 200 University Avenue, Waterloo, ON N2L 3G1 (Canada)
Using a Varian 21 EX linear accelerator with a multileaf collimator (MLC) of 120 leaves, the penumbra regions of beam profiles within an irregular multileaf collimated fields were studied. MLC fields with different leaf stepping angles from 21.8 deg. to 68.2 deg. were used. Beam profiles in different directions: (1) along the cross-line and in-line axis (2) along the leaf stepping edges of the field, and (3) parallel to the stepping edges but in the middle of the field, were measured and calculated using Kodak XV radiographic film and Pinnacle3 treatment planning system version 7.4f. These beam profiles were measured and calculated at source to axis distance=100 cm with 5 cm of solid water slab on top. On the one hand, for both cross-line and in-line beam profiles, the penumbra widths of 20%-80% did not vary with the leaf stepping angles and were about 0.4 cm. On the other hand, the penumbra widths of 10%-90% of the above two profiles varied with the stepping angles and had maximum widths of about 1.9 cm (cross-line) and 1.65 cm (in-line) for stepping angles of 38.7 deg. and 51.3 deg., respectively. For profiles crossing the 'rippled' stepping edges of the field, the penumbra widths (10%-90%) at the regions between two opposite leaves (i.e., profile end at the Y1/Y2 jaw position) decreased with the stepping angles. At the penumbra regions between two leaf edges with the tongue-and-groove structure of the same bank (i.e., profile end at the X1/X2 jaw position), the penumbra widths increased with the stepping angles. When the penumbra widths were measured between two opposite leaf edges and at corners between two leaves, the widths first decreased with the stepping angles and then increased beyond the minimum width point at stepping angle of 45 deg. The penumbra width (10%-90%) measured at the leaf edge was larger than that at the corner. For the beam profiles calculated using Pinnacle3, it is found that the results agreed well with the measurements along the cross-line and in-line axis, while there was a deviation for the profiles along the leaf stepping edge of the field compared to the film measurements. The measured results in this study can help us to understand the dosimetric effect of the leaf stepping (due to finite leaf width), tongue-and-groove and rounded leaf end structure in the penumbra region of an irregular MLC field. A more dedicated penumbra model can be developed for the treatment planning system.
OSTI ID:
20853829
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 12 Vol. 33; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

Characterization of a multileaf collimator system
Journal Article · Thu Jan 14 23:00:00 EST 1993 · International Journal of Radiation Oncology, Biology and Physics; (United States) · OSTI ID:6614872

An MLC calibration method using a detector array
Journal Article · Thu Oct 15 00:00:00 EDT 2009 · Medical Physics · OSTI ID:22102106

Dosimetric characterization of a multileaf collimator for a new four-dimensional image-guided radiotherapy system with a gimbaled x-ray head, MHI-TM2000
Journal Article · Wed Sep 15 00:00:00 EDT 2010 · Medical Physics · OSTI ID:22098527