Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Mass detection sensitivity of piezoelectric cantilevers with a nonpiezoelectric extension

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.2202913· OSTI ID:20853302
; ;  [1]
  1. Department of Materials Science and Engineering, Drexel University, Philadelphia, Pennsylvania 19104 (United States)

A piezoelectric cantilever (PEC) is a transverse transducer consisting of a piezoelectric layer, e.g., lead zirconate titanate (PZT), bonded to a nonpiezoelectric layer, e.g., stainless steel, which has a uniform cross section throughout the length. A PEC with a thin nonpiezoelectric extension, on the other hand, has two distinctive sections each with its own thickness and transverse mass-density and elastic-modulus profiles. A piezoelectric cantilever with a nonpiezoelectric extension has been increasingly used as an in situ biosensor that has the advantage of dipping only the nonpiezoelectric extension part in an aqueous solution without electrically insulating the piezoelectric section. In this study, we examined both experimentally and theoretically the effect of the thin nonpiezoelectric extension, in particular, its length ratio to the piezoelectric part on the vibration wave form and mass detection sensitivity of a PEC. We showed that the nonpiezoelectric extension caused substantial distortion to the vibration wave form. Due to the wave form distortion, the mass detection sensitivity of a PEC with a nonpiezoelectric extension can be higher than that of a PEC of the same length without a nonpiezoelectric extension. Using PECs consisting of a 0.25 mm thick PZT layer and a 0.07 mm thick stainless steel of various lengths, l{sub 1}, and a 0.07 mm thick nonpiezoelectric stainless steel extension of various lengths, l{sub 2}, we showed that for any l{sub 1}, the maximum mass detection sensitivities of first, second, and third modes occurred at l{sub 2}/l{sub 1}=0.6, 0.28, and 0.16 and the maximum values were, respectively, 1.42, 3.9, and 6.7 times the mass detection sensitivity of a PEC of the same l{sub 1} without a nonpiezoelectric extension.

OSTI ID:
20853302
Journal Information:
Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 6 Vol. 77; ISSN 0034-6748; ISSN RSINAK
Country of Publication:
United States
Language:
English

Similar Records

Transverse piezoelectric coefficient measurement of flexible lead zirconate titanate thin films
Journal Article · Thu May 28 00:00:00 EDT 2015 · Journal of Applied Physics · OSTI ID:22410248

Novel high vacuum scanning force microscope using a piezoelectric cantilever and the phase detection method
Journal Article · Tue Jul 01 00:00:00 EDT 1997 · Journal of Vacuum Science and Technology. B, Microelectronics Processing and Phenomena · OSTI ID:562932

Precise measurement of the transverse piezoelectric coefficient for thin films on anisotropic substrate
Journal Article · Sun Jan 27 23:00:00 EST 2013 · Journal of Applied Physics · OSTI ID:22102243