Title: Automated gamma knife radiosurgery treatment planning with image registration, data-mining, and Nelder-Mead simplex optimization

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.2207314· OSTI ID:20853229
; ;  [1]
  1. Unit of Academic Radiology, University of Sheffield, Sheffield, South Yorkshire (United Kingdom)

Gamma knife treatments are usually planned manually, requiring much expertise and time. We describe a new, fully automatic method of treatment planning. The treatment volume to be planned is first compared with a database of past treatments to find volumes closely matching in size and shape. The treatment parameters of the closest matches are used as starting points for the new treatment plan. Further optimization is performed with the Nelder-Mead simplex method: the coordinates and weight of the isocenters are allowed to vary until a maximally conformal plan specific to the new treatment volume is found. The method was tested on a randomly selected set of 10 acoustic neuromas and 10 meningiomas. Typically, matching a new volume took under 30 seconds. The time for simplex optimization, on a 3 GHz Xeon processor, ranged from under a minute for small volumes (<1000 cubic mm, 2-3 isocenters), to several tens of hours for large volumes (>30 000 cubic mm,>20 isocenters). In 8/10 acoustic neuromas and 8/10 meningiomas, the automatic method found plans with conformation number equal or better than that of the manual plan. In 4/10 acoustic neuromas and 5/10 meningiomas, both overtreatment and undertreatment ratios were equal or better in automated plans. In conclusion, data-mining of past treatments can be used to derive starting parameters for treatment planning. These parameters can then be computer optimized to give good plans automatically.

OSTI ID:
20853229
Journal Information:
Medical Physics, Vol. 33, Issue 7; Other Information: DOI: 10.1118/1.2207314; (c) 2006 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English