Artificial heart muscle using SMA wire is developed to assist weaken heartbeat. Simple structure design was adopted for large output force, large displacement and rapid cyclic motion of the actuator. The actuator was designed and fabricated from the viewpoint of heat transfer. Moving performance of the actuator was experimentally measured under 10N of loading condition. Under the maximum efficiency condition, the actuator shows cyclic motion with 1mm of displacement and time period of about 2 seconds in one cycle.