Absolute measurement of F2-laser power at 157 nm
We report a comparison of laser power measurements at the F2-laser wavelength oaf nm made at two facilities of the Physikalisch-Technische Bundesanstalt (PTB), the German national metrology institute. At the PTB laboratory at the electron storage ring BESSY II in Berlin, the scale for laser power was directly traced to a cryogenic radiometer operating at 157 nm, whereas at the PTB laser radiometry facility in Braunschweig the calibration of transfer detectors was performed with a newly developed standard for laser power at 157 nm, which is traceable in several steps to a cryogenic radiometer operating at 633 nm. The comparison was performed under vacuum conditions with laser pulse energies of?10 {mu}J, however with different average powers because different primary standard radiometers were used. The relative deviation for the responsivity of the transfer detector was 4.8% and thus within the combined standard uncertainty.
- OSTI ID:
- 20779332
- Journal Information:
- Applied Optics, Journal Name: Applied Optics Journal Issue: 14 Vol. 45; ISSN 0003-6935; ISSN APOPAI
- Country of Publication:
- United States
- Language:
- English
Similar Records
Gas-Monitor Detector for Intense and Pulsed VUV/EUV Free-Electron Laser Radiation
A new facility for the synchrotron radiation-based calibration of transfer radiation sources in the ultraviolet and vacuum ultraviolet spectral range