Effect of Te precipitates on the performance of CdZnTe detectors
- Department of Nonproliferation and National Security, Brookhaven National Laboratory, Upton, New York 11973-5000 (United States)
Measurements using the National Synchrotron Light Source provided a detailed comparisons of the microscale detector response and infrared microscopy images for CdZnTe Frisch-ring x-ray and gamma detectors. Analysis of the data showed conclusively that local deteriorations of the electron charge collection and x-ray device response fully correlate with the presence of Te precipitates as seen in the IR images. Effects of the surface processing conditions on the detector performance were also clearly observed.
- OSTI ID:
- 20779170
- Journal Information:
- Applied Physics Letters, Journal Name: Applied Physics Letters Journal Issue: 14 Vol. 88; ISSN APPLAB; ISSN 0003-6951
- Country of Publication:
- United States
- Language:
- English
Similar Records
EFFECT OF TE PRECIPITATES ON THE PERFORMANCE OF CDZNTE DETECTORS.
TE INCLUSIONS AND THEIR RELATIONSHIP TO THE PERFORMANCE OF CDZNTE DETECTORS.
Performance Characteristics of Frisch-Ring CdZnTe Detectors
Conference
·
Tue Sep 20 00:00:00 EDT 2005
·
OSTI ID:860508
TE INCLUSIONS AND THEIR RELATIONSHIP TO THE PERFORMANCE OF CDZNTE DETECTORS.
Conference
·
Sun Aug 13 00:00:00 EDT 2006
·
OSTI ID:893015
Performance Characteristics of Frisch-Ring CdZnTe Detectors
Journal Article
·
Sat Dec 31 23:00:00 EST 2005
· IEEE Trans. Nucl. Sci.
·
OSTI ID:914132