Performance evaluation of a computed radiography imaging device using a typical 'front side' and novel 'dual side' readout storage phosphors
- Diagnostic Radiology Department, Mayo Clinic, Rochester, Minnesota 55905 (United States)
The Fourier-space modulation transfer function (MTF), normalized noise power spectrum (nNPS), and detective quantum efficiency (DQE) of a computed radiography (CR) x-ray imaging device were measured. Two different storage phosphor screens were used in conjunction with a single CR reader (Fuji, Clearview, CS). One of the storage phosphor plates (ST-BD) had a clear backing material which allowed 'dual side read' of the latent image from both the 'front' and 'back' sides of the phosphor. The other phosphor plate had a light occluding backing material, limiting the readout to front side only (ST-VI). The standard RQA-5 beam quality was used. The MTF was measured using a 1 mm thick tungsten edge device. Compared to the ST-55BD phosphor, the ST-VI phosphor was found to have modestly higher MTF at all spatial frequencies. The nNPS(f) and DQE(f) were measured for nominal incident exposure levels ranging from 0.1 to 10 mR. The dual side read phosphor demonstrated superior DQE, especially at low spatial frequencies. At the frequency 0.5 cycles/mm, the DQE values for the 1 mR exposure were 0.36 and 0.21 for the ST-55BD and ST-VI phosphor plates, respectively. The differences between the spatial-frequency dependent DQE of the two plates can be attributed to the increased signal collection efficiency of the dual side read plates and differences in storage phosphor structure noise.
- OSTI ID:
- 20775052
- Journal Information:
- Medical Physics, Vol. 33, Issue 2; Other Information: DOI: 10.1118/1.2143306; (c) 2006 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
- Country of Publication:
- United States
- Language:
- English
Similar Records
Physical evaluation of a needle photostimulable phosphor based CR mammography system
Influence of cassette type on the DQE of CR systems