Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Feasibility of a fast inverse dose optimization algorithm for IMRT via matrix inversion without negative beamlet intensities

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.2030427· OSTI ID:20726249
; ;  [1]
  1. Department of Physics and Astronomy, University of Western Ontario, London, Ontario N6A 3K7 (Canada)
A fast optimization algorithm is very important for inverse planning of intensity modulated radiation therapy (IMRT), and for adaptive radiotherapy of the future. Conventional numerical search algorithms such as the conjugate gradient search, with positive beam weight constraints, generally require numerous iterations and may produce suboptimal dose results due to trapping in local minima. A direct solution of the inverse problem using conventional quadratic objective functions without positive beam constraints is more efficient but will result in unrealistic negative beam weights. We present here a direct solution of the inverse problem that does not yield unphysical negative beam weights. The objective function for the optimization of a large number of beamlets is reformulated such that the optimization problem is reduced to a linear set of equations. The optimal set of intensities is found through a matrix inversion, and negative beamlet intensities are avoided without the need for externally imposed ad-hoc constraints. The method has been demonstrated with a test phantom and a few clinical radiotherapy cases, using primary dose calculations. We achieve highly conformal primary dose distributions with very rapid optimization times. Typical optimization times for a single anatomical slice (two dimensional) (head and neck) using a LAPACK matrix inversion routine in a single processor desktop computer, are: 0.03 s for 500 beamlets; 0.28 s for 1000 beamlets; 3.1 s for 2000 beamlets; and 12 s for 3000 beamlets. Clinical implementation will require the additional time of a one-time precomputation of scattered radiation for all beamlets, but will not impact the optimization speed. In conclusion, the new method provides a fast and robust technique to find a global minimum that yields excellent results for the inverse planning of IMRT.
OSTI ID:
20726249
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 9 Vol. 32; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

Real-time fast inverse dose optimization for image guided adaptive radiation therapy-Enhancements to fast inverse dose optimization (FIDO)
Journal Article · Fri May 15 00:00:00 EDT 2009 · Journal of Applied Physics · OSTI ID:21352199

Effect of beamlet step-size on IMRT plan quality
Journal Article · Mon Nov 14 23:00:00 EST 2005 · Medical Physics · OSTI ID:20726909

Dose-volume based ranking of incident beam direction and its utility in facilitating IMRT beam placement
Journal Article · Sat Oct 01 00:00:00 EDT 2005 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:20702169