Verification techniques and dose distribution for computed tomographic planned supine craniospinal radiation therapy

Journal Article · · Medical Dosimetry

A modified 3-field technique was designed with opposed cranial fields and a single spinal field encompassing the entire spinal axis. Two methods of plan verifications were performed before the first treatment. First, a system of orthogonal rulers plus the thermoplastic head holder was used to visualize the light fields at the craniospinal junction. Second, film phantom measurements were taken to visualize the gap between the fields at the level of the spinal cord. Treatment verification entailed use of a posterior-anterior (PA) portal film and placement of radiopaque wire on the inferior border of the cranial field. More rigorous verification required a custom-fabricated orthogonal film holder. The isocenter positions of both fields when they matched were recorded using a record-and-verify system. A single extended distance spinal field collimated at 42 degree sign encompassed the entire spinal neuraxis. Data were collected from 40 fractions of craniospinal irradiation (CSI). The systematic error observed for the actual daily treatments was -0.5 mm (underlap), while the stochastic error was represented by a standard deviation of 5.39 mm. Measured data across the gapped craniospinal junction with junction shifts included revealed a dose ranging from 89.3% to 108%. CSI can be performed without direct visualization of the craniospinal junction by using the verification methods described. While the use of rigorous film verification for supine technique may have reduced the systematic error, the inability to visualize the supine craniospinal junction on skin appears to have increased the stochastic error compared to published data on such errors associated with prone craniospinal irradiation.

OSTI ID:
20655449
Journal Information:
Medical Dosimetry, Journal Name: Medical Dosimetry Journal Issue: 2 Vol. 28; ISSN 0958-3947; ISSN MEDOEJ
Country of Publication:
United States
Language:
English

Similar Records

SU-E-T-255: Optimized Supine Craniospinal Irradiation with Image-Guided and Field Matched Beams
Journal Article · Sun Jun 01 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22351078

A Simple Technique of Supine Craniospinal Irradiation
Journal Article · Tue Apr 01 00:00:00 EDT 2008 · Medical Dosimetry · OSTI ID:21045990

Conventional Craniospinal Irradiation with Patient Supine and Source-Skin Distance (SSD) 100 cm for Spinal Field
Journal Article · Fri Dec 31 23:00:00 EST 2010 · Medical Dosimetry · OSTI ID:22130355