Title: Small prototype of Anger camera with submillimeter position resolution

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.1891485· OSTI ID:20646480
;  [1];  [2]
  1. Dipartimento di Elettronica e Informazione, Politecnico di Milano, Piazza Leonardo da Vinci 32, Milan 20133 (Italy)
  2. Italy

In this work we present a small prototype of Anger camera based on a monolithic array of silicon drift detectors (SDDs) used as photodetectors for a single CsI(Tl) scintillator crystal. This prototype, having a total sensitive area of about 1 cm{sup 2}, has been realized in order to evaluate the performances attainable in {gamma}-ray imaging by using a SDD array instead of the more conventionally employed photomultiplier tubes. An intrinsic resolution better than 200 {mu}m has been measured with this camera. This result was achieved thanks to the extremely low electronics noise presented by the SDD units, fully exploited by the integration of the front-end junction field effect transistor directly on the detector chip. The main features of the detector as well as the results of its extended experimental characterization with 60 and 122 keV gamma rays are presented and discussed in the article. The present device has many potential applications in the field of medical imaging, in which outstanding position resolutions are essential.

OSTI ID:
20646480
Journal Information:
Review of Scientific Instruments, Vol. 76, Issue 4; Other Information: DOI: 10.1063/1.1891485; (c) 2005 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English

Similar Records

The DRAGO gamma camera
Journal Article · Thu Apr 15 00:00:00 EDT 2010 · Review of Scientific Instruments · OSTI ID:20646480

Sub-millimeter resolution SiPM-based neutron anger camera
Journal Article · Fri Nov 03 00:00:00 EDT 2023 · Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment · OSTI ID:20646480

A thick anger camera for gamma-ray astronomy
Conference · Fri Feb 01 00:00:00 EST 1985 · IEEE Trans. Nucl. Sci.; (United States) · OSTI ID:20646480