Preliminary performance measurements for a streak camera with a large-format direct-coupled charge-coupled device readout
Journal Article
·
· Review of Scientific Instruments
- Lawrence Livermore National Laboratory, Box 808, Livermore, California 94551 (United States)
The University of Rochester's Laboratory for Laser Energetics (Rochester, New York) is leading an effort to develop a modern, fully automated streak camera. Characterization of a prototype camera shows spatial resolution better than 20 lp/mm, temporal resolution of 12 ps, line-spread function of 40 {mu}m (full width at half maximum) contrast transfer ratio of 60% at 10 lp/mm, system gain of 101 charge-coupled device electrons per photoelectron, and a dynamic range of 500 for a 2 ns window.
- OSTI ID:
- 20641226
- Journal Information:
- Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 10 Vol. 75; ISSN 0034-6748; ISSN RSINAK
- Country of Publication:
- United States
- Language:
- English
Similar Records
Preliminary Performance Measurements for a Streak Camera with a Large-Format Direct-Coupled CCD Readout
Streak Camera Performance with Large-Format CCD Readout
Direct readout devices for streak cameras
Conference
·
Tue Apr 13 00:00:00 EDT 2004
·
OSTI ID:15014094
Streak Camera Performance with Large-Format CCD Readout
Conference
·
Tue Jul 08 00:00:00 EDT 2003
·
OSTI ID:15005836
Direct readout devices for streak cameras
Conference
·
Mon Aug 16 00:00:00 EDT 1976
·
OSTI ID:7348680