skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Grazing-incidence mirror streak camera diagnostic for emission measurements of imploding z pinches on the Sandia Z-machine

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.1787933· OSTI ID:20641209
; ; ; ; ; ; ;  [1]
  1. Sandia National Laboratories, P.O. Box 5800, MS 1193, Albuquerque, New Mexico 87185 (United States)

A soft x-ray (0.1-1 keV) streak camera using a grazing-incidence mirror has been developed for the Sandia Z facility, a 20 MA, 100 ns rise-time accelerator that can generate >200 TW,2 MJ, x-ray pulses. The streak camera is used to measure with one dimension of spatial resolution the continuous time history of sub-kilo-electron-volts emission from z-pinch and radiation flow experiments. Radiation >1 keV is eliminated by the use of a grazing-incidence mirror and transmission filters. The diagnostic has a magnification of 1.22, a 20 mm field-of-view and a spatial resolution of {approx}350 {mu}m. The recording length of the instrument is variable in duration up to a maximum of 33 ns, making it well suited for the typical 100 ns Z pinch implosion times. The spatial resolution can readily be changed to be along either the axial or the radial direction.

OSTI ID:
20641209
Journal Information:
Review of Scientific Instruments, Vol. 75, Issue 10; Conference: 15. topical conference on high temperature plasma diagnostics, San Diego, CA (United States), 19-22 Apr 2004; Other Information: DOI: 10.1063/1.1787933; (c) 2004 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English