Research and development of x-ray imaging crystal spectrometers for KSTAR
Journal Article
·
· Review of Scientific Instruments
- Korea Basic Science Institute, Taejeon 305-333 (Korea, Republic of)
The engineering design for two high-resolution x-ray imaging crystal spectrometers for KSTAR tokamak has been completed. A spherically bent quartz crystal and a large area two-dimensional (2D) position-sensitive multiwire proportional counter have been selected for the imaging spectrometers. A prototype 10 cm by 30 cm 2D detector was fabricated and pilot measurement on Alcator C-Mod tokamak was carried out. The final engineering design of the spectrometers and experimental results from the 2D detector are presented.
- OSTI ID:
- 20636723
- Journal Information:
- Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 10 Vol. 75; ISSN 0034-6748; ISSN RSINAK
- Country of Publication:
- United States
- Language:
- English
Similar Records
Development of advanced x-ray imaging crystal spectrometer utilizing a large area segmented proportional counter for KSTAR
Spatially Resolved Spectra from a new X-ray Imaging Crystal Spectrometer for Measurements of Ion and Electron Temperature Profiles
Spatially resolved spectra from a new x-ray imaging crystal spectrometer for measurements of ion and electron temperature profiles (invited)
Journal Article
·
Fri Jun 15 00:00:00 EDT 2007
· Review of Scientific Instruments
·
OSTI ID:20953468
Spatially Resolved Spectra from a new X-ray Imaging Crystal Spectrometer for Measurements of Ion and Electron Temperature Profiles
Journal Article
·
Tue Aug 10 00:00:00 EDT 2004
· Review of Scientific Instruments
·
OSTI ID:15014632
Spatially resolved spectra from a new x-ray imaging crystal spectrometer for measurements of ion and electron temperature profiles (invited)
Journal Article
·
Fri Oct 01 00:00:00 EDT 2004
· Review of Scientific Instruments
·
OSTI ID:20636715