Validation of Monte Carlo calculated surface doses for megavoltage photon beams
- Department of Medical Physics, McGill University Health Centre, 1650 Avenue Cedar, Montreal, Quebec, H3G 1A4 (Canada)
Recent work has shown that there is significant uncertainty in measuring build-up doses in megavoltage photon beams especially at high energies. In this present investigation we used a phantom-embedded extrapolation chamber (PEEC) made of Solid Water{sup TM} to validate Monte Carlo (MC)-calculated doses in the dose build-up region for 6 and 18 MV x-ray beams. The study showed that the percentage depth ionizations (PDIs) obtained from measurements are higher than the percentage depth doses (PDDs) obtained with Monte Carlo techniques. To validate the MC-calculated PDDs, the design of the PEEC was incorporated into the simulations. While the MC-calculated and measured PDIs in the dose build-up region agree with one another for the 6 MV beam, a non-negligible difference is observed for the 18 MV x-ray beam. A number of experiments and theoretical studies of various possible effects that could be the source of this discrepancy were performed. The contribution of contaminating neutrons and protons to the build-up dose region in the 18 MV x-ray beam is negligible. Moreover, the MC calculations using the XCOM photon cross-section database and the NIST bremsstrahlung differential cross section do not explain the discrepancy between the MC calculations and measurement in the dose build-up region for the 18 MV. A simple incorporation of triplet production events into the MC dose calculation increases the calculated doses in the build-up region but does not fully account for the discrepancy between measurement and calculations for the 18 MV x-ray beam.
- OSTI ID:
- 20634631
- Journal Information:
- Medical Physics, Journal Name: Medical Physics Journal Issue: 1 Vol. 32; ISSN 0094-2405; ISSN MPHYA6
- Country of Publication:
- United States
- Language:
- English
Similar Records
Build-up and surface dose measurements on phantoms using micro-MOSFET in 6 and 10 MV x-ray beams and comparisons with Monte Carlo calculations
SU-F-T-371: Development of a Linac Monte Carlo Model to Calculate Surface Dose
Radiation induced currents in parallel plate ionization chambers: Measurement and Monte Carlo simulation for megavoltage photon and electron beams
Journal Article
·
Sun Apr 15 00:00:00 EDT 2007
· Medical Physics
·
OSTI ID:20951147
SU-F-T-371: Development of a Linac Monte Carlo Model to Calculate Surface Dose
Journal Article
·
Wed Jun 15 00:00:00 EDT 2016
· Medical Physics
·
OSTI ID:22648969
Radiation induced currents in parallel plate ionization chambers: Measurement and Monte Carlo simulation for megavoltage photon and electron beams
Journal Article
·
Fri Sep 15 00:00:00 EDT 2006
· Medical Physics
·
OSTI ID:20853428